Single Lab Validation of a Method for the Determination of β‐

Galactooligosaccharides in Infant Formula & Adult Nutritionals / Sean Austin (NRC, 

Lausanne)  

25 Jul 2016

CONFIDENTIAL

 ∙ This document may not be reproduced or disclosed to third parties without prior authorization 

Page 10 / 35

2.2.6

Chromatographic Conditions 

Analyse the prepared samples and standards on a UHPLC system equipped with a FLD under the 

following conditions: 

FLD: Excitation λ: 330 nm   Emission λ: 420 nm 

Trap Column:  Acquity BEH Amide (2.1 mm x 5 mm; 1.7 µm) or equivalent (see Figure 1 for 

plumbing) 

Column: Acquity UPLC BEH Glycan (2.1 mm x 150 mm; 1.7 µm) or equivalent 

Column temperature: 25°C ± 1 °C 

Injection volume: 3 µL 

Eluent A: Acetonitrile 

Eluent B: Ammonium formate 0.1 mol/L, pH 4.4 

Gradient: 

Time 

(min) 

Flow 

(mL/min) 

%A 

(Acetonitrile) 

%B 

(Formate) 

Switching 

Valve 

Comment 

0.6 

97 

6‐1 

Inject sample on trap column 

5.0 

0.6 

97 

6‐1 

End of sample wash, start acquisition 

5.5 

0.6 

88 

12 

1‐2 

Switch valve, start analytical gradient 

12.5 

0.6 

88 

12 

1‐2 

22.5 

0.6 

83 

17 

1‐2 

26.5 

0.6 

83 

17 

1‐2 

45.0 

0.6 

68 

32 

1‐2 

End of analytical gradient 

45.1 

0.2 

68 

32 

1‐2 

Reduce flow 

45.5 

0.2 

30 

70 

1‐2 

Start column wash 

55.5 

0.2 

30 

70 

1‐2 

End column wash 

57.0 

0.2 

88 

12 

1‐2 

Start analytical column equilibration 

62.0 

0.2 

88 

12 

1‐2 

72.0 

0.6 

88 

12 

1‐2 

End analytical column equilibration 

72.5 

0.6 

97 

6‐1 

Switch valve, equilibrate trap column 

73.5 

0.6 

97 

6‐1 

End 

Figure 1: Configuration of 2-way 6-port switching valve