Results
3D-CRT Step and
shoot
Sliding
window
RapidArc
PTV
EVAL
V95% 99.7
99.7
99.7
99.6
PTV
EVAL
COIN 0.62
0.85
0.85
0.86
Ipsilat. breast
V50% 44.1
36.3
35.6
29.4
Ipsilat. lung
V10% 10.1
34.3
34.3
35.3
Heart (left)
V5% 1.5
22.6
20.9
48.3
Contralat. breast V5% 0
8.1
8.3
3.1
Contralat. lung V5% 0
8.4
8.3
6.3
-
good target volume
coverage can be achieved with all four techniques
-
conformality
is better with the
IMRT
techniques
-
ipsilateral breast can be better protected with IMRT
techniques (the best is the RapidArc)
-
contralateral breast and lung receive higher dose with IMRT
techniques than with 3D-CRT