Sensitivity calibration & energy response:
,
Our calibration coeff. (inverse of dosimeter A.D. sensitivity)
actuall comprises 2 parts
y
:
) (
) (
) (
) ( det
) ( det
) (
) (
) (
) ( ,
QbqkQf
QM
Q D
Q D
QwD
QM
QwD
QwADN
•
we define the
absorbed dose energy dependence
of the detector as the ratio
of dose to water per unit dose to the detector at a given beam quality,
This depends on physical properties and radiation quality (cross
) (
) (
) (
det
Q D
QD Qf
w
sections) and detector geometry.
It can be calculated via MC!
•
We define the
intrinsic energy dependence
of the detector as the ratio of dose
required to be absorbed to produce a unit signal
) (
) (
det
Q D Q k
b
This depends mainly on the physical process underlying the conversion of
dose to the measured signal (and hence LET).
It can only be measured!
) (
QM
q