Brachy dosimetry is challenging …

f

zyD

) (

.

or an Ir-192 HDR source

K

S

,

Dose rate varies from ~20 Gy/s @

(0.5cm,90

0

) from an Ir-192 HDR source

to ~2 μGy/s @ (5cm,90

0

) from an I-125

LDR source

Sensitivity high for low uncertainty dose ,

range high, (preferably) with k

l

=k

dr

=1

Spatial dose gradient is high (i.e.

~25%/mm in the radial direction @

(0.5cm,90

0

) from an Ir-192 HDR source)

Volume must be small

positional

,

accuracy is very important

(1% rel. uncert. in D(1cm,90

0

) requires

0.05mm uncertainty in r, assuming 1/r

2

dose dependence since:

)

If solid phantoms are used to increase

r

D

r

D

2

positional accuracy a correction is

required from D

w,phant

to D

w,w

at each

meas position

.

Figure from: Taylor, Yegin, Rogers, Med. Phys. 34(2) 445 (2007)