TLD: sensitivity/linearity

i

ff i

i

f i

i

i h

TLD s e ect ve n terms o nteract ng w t

photons in the brachytherapy E range

The TL mechanism however is inefficient

About 0.04% of TLD absorbed dose is emitted

as TL energy per unit mass

Individual TLD calibration is required as well as

meticulous care in reproducible conditions of

use to ensure precision/accuracy

TLD dynamic dose range is wide. It comprises a

linear D region followed by a region of supra-

linearity and eventually saturation

,

,

Linearity cannot be assumed and has to be

measured by irradiating TLD groups in graded

doses in the region of interest

Table from: Mayles, Nahum, Rosenwald (Eds): Handbook of radiotherapy physics: theory and practice,©

2007 by Taylor & Francis Group LLC