Table of Contents Table of Contents
Previous Page  132 / 258 Next Page
Information
Show Menu
Previous Page 132 / 258 Next Page
Page Background

1406 

Jing

et al.

: J

ournal of

AOAC I

nternational

Vol. 98, No. 5, 2015

at the primary transition varied on a given day and a given

instrument. Table 12 shows the results.

Daily ratios for carnitine were, overall, more consistent

than those for choline, especially for data collected on the

TQS instrument. For 4 of the 12 days of TQS choline data,

the transition ratio was approximately twice that observed on

the remaining days. There is no obvious explanation for this

shift. Within-day uniformity was generally good on those days

(within an acceptable range), although on 3 of those 4 days the

ratio RSDs were modestly larger. Despite the large shift in the

ion ratio on those days, however, there was no other indication

that data quality was compromised. Overall, within-day

RSD ranges were carnitine/TQD: 1.3–2.7%, carnitine/

TQS: 0.6–3.6%, choline/TQD: 0.7–1.5%, and choline/TQS:

1.1–5.4%. These are consistent with expectations, generally, for

LC/MS/MS methods.

Because of the differences between instrument platforms as

well as day-to-day shifts, it is not possible to assign meaningful

ranges appropriate for all instruments on all days. A better

alternative is to base the expected ratio in the samples to that

determined for the current daily calibration standards, and to

use some multiple of the SD of the ratio determined for the three

standards to set a limit for the ratio expected in the samples.

Thus, if a particular sample’s ratio was 2–3 SD away (the exact

setting is arbitrary) from the typical ratio in the daily calibration

standards, that would be flagged as a possibly invalid result

(caused by a coeluting molecule that happened to have exactly

the same parent and primary daughter ion mass) and investigated

(e.g., diluted and reanalyzed). The AOAC ERP has previously

suggested this type of approach for MS/MS platforms to deliver

the most highly specific methods to the analytical community.

As reasonable as this approach sounds, we have found it very

difficult to put into practice for LC/MS/MS methods. The flags

raised can disrupt laboratory operations, and the subsequent

investigation/retest is often ambiguous as to whether the result

is more accurate, or even statistically different from the flagged

result. Given the excellent performance of AOAC

2014.04

and good agreement among different methods on a variety of

matrixes, we conclude that confirmatory ion transitions are not

necessary for this method.

Conclusions

An SLV of AOAC

2014.04

has been performed, and the

method has met all the requirements for precision, accuracy,

linearity, LOQ, and specificity as outlined in SMPR

2012.010

for carnitine. The method simultaneously determines choline,

and the data show that it has also met the requirements of SMPR

2012.013

for that analyte.

References

 (1) Starkey, D.E., Denison, J.E., Seipelt, C.T., & Jacobs, W.A.

(2008) J. AOAC Int . 91 , 130–142

 (2) Andrieux, P., Kilinc, T., Perrin, C., & Campos-Giménez, E.,

(2008) J. AOAC Int . 91 , 777–785

 (3) Fu, S., Tao, B., Lai, S., Zhang, J., & Yiping, R.

(2012) J. AOAC Int . 95 , 157–162

 (4) Phillips, M.P., & Sander L.C

. (2012) J. AOAC Int . 95 , 1479–1486

 (5) Oates, K., Chen, L., DeBorba, B., Rohrer, J., Mohindra, D.,

Thermo Fisher Scientific Application Note 193.

www.dionex. com/en-us/webdocs/115182-AU193-IC-Choline-InfantFormula- AdultNutritional-AU70918_E.pdf

(accessed August 9, 2015)

132