Table of Contents Table of Contents
Previous Page  138 / 258 Next Page
Information
Show Menu
Previous Page 138 / 258 Next Page
Page Background

1660

Butler-Thompson et al.:

J

ournal of

AOAC I

nternational

V

ol.

98, N

o.

6, 2015

laboratory water and discard eluent. Air-dry each cartridge by

pulling a vacuum until no more effluent is observed. Close

each stopcock. Place a 5 or 10 mL volumetric flask under each

cartridge. Add 4.4 mL 30% acetonitrile to each cartridge. Open

each stopcock and elute vitamin B

12

into the volumetric flasks.

(

4

) 

Final dilution

.—For samples collected in 10 mL

volumetric flasks, dilute to volume with water. For samples

collected in 5 mL volumetric flasks, in a hood add 0.1 mL

freshly prepared 0.4% KCN to each volumetric flask. Place

prepared samples in a 95°C oven for at least 1.5 h, but for no

more than 4 h. After at least 1.5 h, remove samples from the oven

and cool to room temperature. Dilute to volume with laboratory

water. Filter an aliquot of each standard and prepared sample

through a 0.45 µm syringe filter into an autosampler vial.

(c)

HPLC analysis

.—(

1

) 

System setup and configuration.—

See

Figures

2011.10A

and

B

for configurations.

(

2

) 

Instrument

operation

conditions

.—(

a

) 

Run

time

.—30–35 min.

(

b

) 

Injection volume

.—900 µL to 2.0 mL.

(

c

) 

System configuration.—See

Table

2011.10E

.

(

d

) 

Isocratic pump

.—Mobile phase D: 2.5% acetonitrile.

Flow rate: Adjust so that vitamin B

12

elutes from the size-

exclusion column between 10.5 and 14.5 min. Typical flow

rates, 1.1–1.2 mL/min.

Note

: To determine an appropriate

flow rate, connect the size-exclusion column directly to the

UV-Vis detector and inject the high standard. Adjust flow rate as

necessary so that vitamin B

12

elutes between 10.5 and 14.5 min.

(

e

) 

Gradient pump

.—Mobile phase compositions: mobile

phaseA, 0.4%TEAin laboratory water, pH 5–7; mobile phase B,

0.4% TEA and 25% acetonitrile in H

2

O, pH 5–7; mobile phase

C, 0.4% TEA and 75% acetonitrile in H

2

O, pH 5–7. Determine

an appropriate gradient to elute vitamin B

12

in 23–30 min and

resolve vitamin B

12

from riboflavin using the information in

Table

2011.10F

. (

See

Figure

2011.10C

.)

(

f

)

Gradient pump flow rate

.—1.0 mL/min.

(

g

) 

Detector

settings

.—Detection

wavelengths

and

bandwidth, 550 and 10 nm, respectively.

(

3

) 

HPLC of standards and samples

.—Make 3–4 injections

of a working standard and verify the precision of those injections

is ≤3%. If the system is working properly, inject a set of 3–6

working standards once, a set of 1–14 samples, and another set

of 3–6 working standards. Every set of 1–14 samples should be

bracketed by standards of appropriate concentration.

F. Calculations

(a) 

Chromatography

.—Visually inspect each standard and

sample chromatogram and verify that vitamin B

12

is resolved

from all other peaks in the chromatograms (Figures

2011.10D

and

E

).

(b) 

Measurement of peak area

.—Peak areas are measured

with a data system. Before calculating the vitamin B

12

concentrations of samples, compare the vitamin B

12

peak areas

of the standards with the vitamin B

12

peak areas of the samples

and verify that the vitamin B

12

peak areas of the samples are

within the range of the vitamin B

12

peak areas of the standards.

(c) 

Calculation of standard concentration

.—

WS = S

w

× P × A/200

where WS = working standard concentration in μg/L;

S

w

 = amount of vitamin B

12

standard weighed in mg;

P = purity of USP reference standard in μg cyanocobalamin

(vitamin  B

12

)/mg of the standard; A = aliquot of vitamin

B

12

intermediate standard used (0.5, 1, 2, 3, 4, or 5) in mL;

and 200 = dilution volume in mL.

(d)

Preparation of standard curves

.—(

1

) At each standard

concentration, average the peak area of the standard injected

at the beginning of a set of samples with the peak area of the

standard injected at the end of the set of samples. Prepare a

standard curve by performing linear least squares regression

Figure 2011.10A. System setup and configuration: Configuration 1.

Figure 2011.10B. System setup and configuration: Configuration 2.

Table 2011.10E. System configuration

Time, min

Valve configuration

0.00–10.5

Configuration 1

10.5–14.5

Configuration 2

14.5–30.0 to 33

Configuration 1

138