2

Not specified

Accuracy/Recovery:

Spike recovery not specified.

Recovery measured as % bias against NIST 1849a, within acceptable limits

Precision (RSD

r

):

Mostly meets SMPR limits for repeatability

Reproducibility (RSD

R

):

Reported, however no context of how reproducibility measured (number labs, etc.). Values

acceptable against SMPR.

System suitability:

No details

Recommendation:

Not recommended First Action at this time as the SPIFAN kit was not used as required by ERP. Also issue of

proprietary nature of analysis should be discussed.