FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

Cons/Weaknesses

Method does not show validation data for any of the SPIFAN matrices (only NIST 1849a)

Methionine sulfone separation has very low peak resolution

Internal standard is added after hydrolysis, transfer and filter steps so it is only adjusting for instrument

recovery and no prep losses.

Supporting Data

General Comment: Method has supporting validation data in NIST SRM 1849a only

-

Method Optimization: Concerns regarding resolution of methionine sulfone peak.

Performance Characteristics:

Analytical Range: 3.6 mg/100g-22,500mg/100g. (SMPR: 0.4 mg/100g-2500 mg/100g)

LOQ: 3.6 mg/100g on an as is basis (SMPR: 0.4 mg/100g on reconstituted powder basis)

Accuracy/Recovery: For NIST 1849a recoveries reported ranged from 92.4-107.6 based on result

obtained vs NIST SRM reference value. No spike recovery data reported in any SPIFAN matrices.

Precision (RSD

r

): On NIST SRM 1849a ranged from 0.6-3.2% RSD (not listed for Trp). It is not clear

how many replicates were performed to determine this value.

Reproducibility (RSD

R

): Reported in NIST SRM 1849a to range from 1.4-4.9% RSD(R) however it is

not clear what values were included and/or how this was determined.

System suitability:

Not specified in current documents.

Recommendation:

Request for SLV data in SPIFAN matrices.

Amino-03 Review Forms