2. Does the method contain

system suitability tests or

controls as specified by the

SMPR? If not, please indicate

if there is a need for such

tests or controls and which

ones.

yes

3. Is there information

demonstrating that the

method system suitability

tests and controls as

specified in the SMPR worked

appropriately and as

expected? If no, please

specify.

yes

4. Based on the supporting

information, is the method

written clearly and concisely?

If no, please specify the

needed revisions.

yes - this method is well written

5. Based on the supporting

information, what are the

pros/strengths of the

method?

This method appears to be rugged and performed well during the SLV study.

6. Based on the supporting

information, what are the

cons/weaknesses of the

method?

This method uses post column dervitization with Ninhydrin which requires some unique

instrumentation which may not be available in labs that are not familiar with amino acid

analysis.

7. Any general comments

about the method?

no

Do you recommend this

method be adopted as a First

Action and published in the

Official Methods of Analysis

of AOAC INTERNATIONAL?

Please specify rationale.

I recommend that this method be considered for AOAC Official First Action status for

the analysis of Theanine.