Commercialised in vivo detectors-diodes

EDP30 EDPHL-3G P30

QED 1116 Isorad-p 1164 Isorad-3 PTW T60010H

Manufacturer

IBA

IBA

ONCOlog

Medical

Sun Nuclear

Sun Nuclear

Sun Nuclear

PTW

Die - type

p

p

n

p

p

n

p

Die – dimensions-area (mm

2

)

1.76

1.76

6.8

0.64

1.54

1.54

1

Die- thickness (μm)

60

60

100

-

-

-

2.5

Build up cap -shape

hemispherical

hemispherical

cylindrical

hemispherical

cylindrical

cylindrical

hemispherical

Build up cap - material

Tantalum Tantalum Tungsten

Brass

Tungsten

Tungsten

Tungsten

Build up cap- water equivalent

thickness (mm)

14

14

30

30.4

26

26

30

Preirradiation dose- modality

8 kGy at 10

MeV

confidential

25 kGy

10 kGy at 10

MeV

10 kGy at 10

MeV

10 kGy at 10

MeV

0 Gy