Table of Contents Table of Contents
Previous Page  1048 / 1359 Next Page
Information
Show Menu
Previous Page 1048 / 1359 Next Page
Page Background

EDP30 EDPHL-3G P30

QED 1116 Isorad-p 1164 Isorad-3 PTW T60010

Stability after irradiation

(5 min)

-0.58% -0.40%

0.33%

-0.06%

-

0.20% 0.43%

0.00%

Intrinsic precission (sd)

(10 irradiations)

0.16% 0.10%

0.05%

0.07%

0.10%

0.64%

0.09%

Lineality response/dose

(r

2

)

1.0000E00

(0.2 Gy-7 Gy)

1.0000E00

(0.2Gy-7 Gy)

1.0000E00

(0.2 Gy-3.5 Gy)

1.0000E0

(0.2 Gy- 7 Gy)

1.0000E00

(0.2 Gy-7 Gy)

1.0000E00

(0.2 Gy-7 Gy)

1.0000E00

(0.2 Gy-2.3 Gy

Depth of the diode measurement point

(water equivalent depth)

1.4 cm 1.4 cm

3.0 cm

2.2 cm

3.3 cm

3.3 cm

3.5 cm

SVWAD

(per 100 Gy)

3.4%

1%

0.2%

0.8%

0.3%

0.3%

1.8%

Dose decrease at 5 cm depth

2.5%

2.5%

10.4%

5.6%

14%

14%

11%

Commercialised in vivo detectors-diodes

Acceptance tests