Are existing definitions of

Tolerance Level

and

Action Level useful

?

If a daily measurement is within the

tolerance level

, no action is required.

If the measured result exceeds the

action level

, immediate action is necessary and

the machine must not be used clinically until the problem is corrected and the

correction is verified by measurement.

However, if the measurement falls

between

tolerance and action levels, this may

be considered acceptable until the next measurement.

If

repeated measurements

remain consistently between tolerance and action

levels, adjustment is required.

Any measurement at any time outside the action level requires immediate

investigation and, if confirmed, rectification.