S113
ESTRO 36 2017
_______________________________________________________________________________________________
OC-0228 DVH criteria for prostate in vivo EPID
dosimetry
R.F.M. Van Oers
1
, E. Van der Bijl
1
, I. Olaciregui-Ruiz
1
, A.
Mans
1
1
Netherlands Cancer Institute, Radiation Oncology,
Amsterdam, The Netherlands
Purpose or Objective
In our department
in vivo
EPID dosimetry is used for dose
verification of all treatment plans. The algorithm uses
EPID images acquired behind the patient to reconstruct
the in vivo 3D dose distribution. This is then automatically
compared to the planned dose distribution and alerts are
generated when deviations are detected.
These alerts are based on γ-analysis. Gamma values
combine dose-difference and distance-to-agreement in a
single metric, but this metric contains no information on
the clinical relevance of deviations. Furthermore, γ-
analysis can be insensitive to systematic under- or
overdoses in plans with inhomogeneous dose distributions.
Dose-volume histograms (DVHs) are widely used for
evaluation of treatment plans, and readily understood by
clinicians. Moreover, differences in DVH parameters can
be linked more straightforwardly to clinical relevance.
This makes DVH-based criteria an attractive alternative to
γ-criteria for in vivo EPID dosimetry alerts.
In this study we investigated the correlation between γ-
and DVH-parameters of the PTV for prostate treatments,
and compared the alert rates for different criteria in order
to propose a suitable set of DVH-criteria for clinical
implementation.
Material and Methods
In vivo 3D dose distributions were reconstructed for the
first three fractions of 95 prostate VMAT treatments, and
then averaged for the evaluation of each treatment. The
γ-analysis was done with global 3%/3mm settings. DVHs
were obtained for the PTV (prostate + seminal vesicles).
Calculated γ-parameters were mean γ, the near-maximum
γ (γ1%), and the γ-passrate (γ%<1); our current criteria
also include the isocenter dose difference (ΔDisoc).
Calculated DVH-parameters were the difference in near-
maximum dose (ΔD2), median dose (ΔD50), and near-
minimum dose (ΔD98). We obtained alert rates for
different sets of criteria on these DVH-parameters. These
were compared to alert rates for the current γ-based
criteria. There are two alert levels, higher-priority “error”
and lower-priority “warning”.
Results
The strongest correlation was found between γ-mean and
|ΔD50|, Pearson’s r=0.95. All other γ- and DVH
parameters were also strongly correlated, with r values
around 0.85.
Figure 1:
Relation between γ-mean and ΔD50 for the
analyzed prostate VMAT plans. The indicated alerts are
generated by the “error level” sets of γ- and DVH-criteria.
Table 1 shows alert rates for different sets of γ- and DVH-
criteria. The two highlighted sets of γ-criteria are the ones
currently used in our clinic to generate alerts for prostate
treatments, juxtaposed with sets of DVH-criteria of similar
alert rate.
Table 1:
Alert rates (% of treatments) for different sets
of γ- and DVH-criteria. The top set of γ-criteria
corresponds to “warning level” alerts, the bottom set
corresponds to “error level” alerts.
Conclusion
A strong correlation was found between γ- and DVH-
parameters of the PTV; a set of DVH-criteria that performs
comparably to the current γ-criteria can easily be chosen.
OC-0229 EPID dose response in the MR-Linac with and
without presence of a magnetic field
I. Torres Xirau
1
, I. Olaciregui-Ruiz
1
, B. J. Mijnheer
1
, U. A.
van der Heide
1
, A. Mans
1
1
Netherlands Cancer Institute Antoni van Leeuwenhoek
Hospital, Department of Radiation Oncology,
Amsterdam, The Netherlands
Purpose or Objective
Image-guided radiotherapy systems are being investigated
and clinically implemented aiming for online and real-time
adaptation of the treatment plan. The use of Electronic
Portal Imaging Devices (EPIDs) for independent in vivo
dose verification in the Elekta MR-Linac is being
developed. One of the challenges for MR-Linac portal
dosimetry is the presence of a small magnetic field at the
EPID level. In the presence of a magnetic field, the
secondary electrons that actually deposit the dose in the
scintillator of the EPID will be affected by the Lorentz
force possibly leading to a B-field induced dose
redistribution. The aim of this study was to analyze and
quantify the effects of the B-field on the EPID images
acquired on the Elekta MR-Linac.
Material and Methods
The Elekta/Philips MR-Linac combines a 1.5T magnetic
resonance imaging scanner with a linear accelerator and
is equipped with an on-board EPID. A magnetometer
(MetroLab THM1176) was used to measure the strength of
the magnetic B-field at the surface of the EPID. To assess
the reproducibility of the panel readouts, a 10x10 cm
2
field was irradiated 10 times in two consecutive days and
the value of the on-axis region (averaged 5x5 pixels) of
EPID images was recorded. During the installation of the
MR-Linac in our institute, EPID images were acquired
before the B-field was ramped up and repeated with B-
field one month later. To study the on-axis response of the
EPID as function of field size with and without the
magnetic B-field, square fields were irradiated with field
sizes varying from 2 to 20 cm. Furthermore, EPID images
acquired with and without B-field were compared by
means of a 2-D γ-analysis (local 2%,1mm, 20% isodose) and
X-Y EPID lateral profiles were compared by visual
inspection.
Results
The magnetic field measured on top of the panel did not
exceed 2.5 mT, yielding an electron trajectory radius of
approximately 1.20 m. The reproducibility of EPID central
axis values for ten irradiated 10x10 cm
2
fields was 0.3% (1
SD).
The normalized on-axis EPID response as function of field
size with and without the presence of the magnetic field
is shown in
Figure 1
together with their ratio.