S108
ESTRO 36 2017
_______________________________________________________________________________________________
medical physicists built a document management system
that follows the patient process from university clinics to
the Skandion Clinic. It contains all guiding and supporting
documents, procedures and instructions. An essential
experience acquired was that it is very important to write
the required documents on so many procedures,
guidelines and instructions as much as possible before the
first patient treatment. Thus it was found that once the
clinic was up and running with patient treatments, it was
much easier to revise an existing procedure that to get
time to write entirely new ones.
Acceptance and measurement, education and training.
It was decided early on that the RTT/nurses would work in
team with the medical physicists on measurement and
acceptance of IBA's equipment so there was at least one
RTT/nurse during the daytime. Then we continued during
the first half of the year with a medical physicist and a
RTT/nurse to work together on the morning checks. The
clinical training included the IBA Web training from UPenn
in Philadelphia as well as study visits for two weeks in the
proton clinic in Philadelphia. To this the IBA on site clinical
training at the Skandion Clinic was added. We were given
training on all new systems including trolley transport, CT
scanning, journal system, fire and safety, CPR, etc.
Another valuable experience was gathered from end-to-
end testing, when a mock patient went through all phases
of the process from each university clinic to the Skandion
Clinic and back.
An empty House to a functioning clinic.
RTT/nurses had responsibility for equipping the premises
with all that may be needed in order to start up a
functioning clinic. This included practically everything
from staff clothing to consumables and medicines, the
toys for the children and even the special requirements of
the anesthesia personnel.
In summary, it has been hard work, frustrating at times,
but extremely educational and stimulating most of the
time, and above all a fun time to start up a clinic.
SP-0217 Workflow in a proton therapy department –
real difference from photon therapy?
F. Fellin
1
1
Ospedale Santa Chiara di Trento, U.O. Protonterapia,
Trento, Italy
In Europe, there was a significantly increase in proton
therapy (PT) facilities in the last few years. In PT,
Radiation Therapy Technologists (RTTs) have a key role
throughout the patient’s therapeutic course, such as in
photon therapy. The physical characteristics of protons
are an advantage for saving the organs at risk (OARs)
and/or for increasing the dose to the target, but they
imply some important criticalities during the patient
workflow. Pencil Beam Scanning (PBS) is the most
advanced technique for PT but this is the proton technique
with more criticality. RTTs, should know these critical
issues and be aware of the impact that they have in the
patient workflow and in their working activities.
RTTs are involved in several steps of the patient workflow
in PT; in general, they are the same as in photon therapy:
simulation CT, diagnostic imaging, treatment planning,
delivery of therapy. However, in each single step, there
are important differences compared to photon therapy.
Starting from the simulation, the main differences relate
to the choice of the immobilization devices, both for their
purchase that for each specific patient, and the definition
with tatoo of the treatment isocenter that, especially with
PBS, it can be defined directly during the acquisition of
simulation CT. In addition, since the dose delivery is very
accurate, the definition of target volumes and OARs must
be very strict. For this reason, the PT facilities are usually
equipped with MR and PET-CT systems; RTTs must be able
to use this equipment and to acquire the necessary
images.
In the treatment planning step, the Medical Dosimetrists
(MDs) work with the Medical Physicists in the plans
optimization; they will use all the experience to get the
best plan for each patient. In addition to all the attention
that MDs must have during treatment planning, which are
common in photon therapy, in PT there are other
situations in which they must be careful. For example, it
is very important to set the correct direction of the fields
based on the tissues that the beams must pass through
before reaching the target and it is recommended to avoid
that all fields have the distal fall-off in the same area (in
particular near OARs); it is important to limit the use of
the range shifter and to select the most appropriate
irradiation technique based on the specific case (SFO,
MFO); MDs must know and evaluate all possible
uncertainties (range, RBE, …); they must adopt robust
optimization techniques and field specific PTV.
Even for the delivery of therapy sessions, there are some
important differences between photon and proton
therapy. First, the equipment is quite different: RTTs
must know in detail the PT equipment operation. Proton
therapy equipment is much larger and more complex
compared to photon therapy system. In PT, the patient
setup is checked every day; generally, RTTs use flat panels
with x-ray tubes system (for 2D images) or CBCT (for 3D
images). In some facilities, a CT on-rails is installed in the
treatment room and RTTs use this one for monitoring the
patient position before the therapy. The advantage of CT
on-rails is that, in principle, you can use the daily images
for checking the dose distribution every day. In PT, tattoos
performed during the simulation CT are important
especially in the first treatment session. They are used to
align the patient and move manually the treatment couch
on the isocenter, for defined the setup position.
Afterwards, RTTs capture the couch coordinate for setup
position and, during subsequent sessions, the table will be
aligned automatically. In all sessions except the first one,
tattoos are important only in treatments with
immobilization devices that are not perfectly indexed on
the couch, e.g. in the case of treatments in pelvic area.
RTTs must be very careful during the proton dose delivery
with PBS. There isn’t a proper ratio between dose and
monitor units delivered; the number of monitor units
depend on the field size: the greater the size of the target,
the greater the number of MU and the beam-on time.
Given the precision of proton dose delivery, small
anatomical changes can lead to important variations of
dose distribution in the patient. For this reason, RTTs must
observe and communicate any anatomical variations that
they see during the patient setup. In this way, it is possible
to perform any dosimetrical checks in the course of
therapy. In these cases, it is very useful to have a CT on-
rails. In PT, RTTs will have to acquire a weekly control CT
for treatments of areas prone to anatomical changes. It
will be used by Physicists or MDs to perform dose
assessments during the treatment and, if necessary, they
will organize a replanning.
These are some of the most significant differences
observed in proton therapy compared to photon therapy;
they change the work of RTTs in patient workflow. PT is
an advanced, complex and precise radiation therapy
technique and it requires very high skills for RTTs.
Symposium with Proffered Papers: Combining tumour
and normal tissue models
SP-0218 Novel approaches in the study of bladder
cancer
A. Kiltie
1
1
Oxford, United Kingdom,
Patients with muscle-invasive bladder cancer can be
treated by removal of their bladder (cystectomy) or by
bladder preserving strategies, which include the use of