Table of Contents Table of Contents
Previous Page  291 / 1082 Next Page
Information
Show Menu
Previous Page 291 / 1082 Next Page
Page Background

S278

ESTRO 36 2017

_______________________________________________________________________________________________

profiles of the individual beams into account was fitted to

the data. The lateral profiles of the WFF beams were

assumed to be uniform while those of the FFF beams were

approximated using 4th or 6th order polynomials. The

scatter functions of the FFF beams were recalculated

using a uniform lateral profile (the same as the physical

profile of the WFF beams) and are henceforth denoted as

virtual uniform FFF beams (VUFFF). The field sizes of the

VUFFF beams having the same scatter contribution as the

corresponding FFF beams at a given field size were defined

as the EQUSFs.

Results

In total, the data of four different LINACs and 18 different

beam energies were analysed. The average values of

EQUSFs over all investigated LINACs of the conventional 10

x 10 cm² reference fields of 6 MV and 10 MV FFF beams for

C-arm LINACs and machine specific reference fields for

CyberKnife and TomoTherapy were 9.5 cm, 9 cm, 5.3 cm

and 6.5 cm, respectively. The standard deviation of these

averaged EQUSFs was below 0.1 cm. Figure 1 illustrates

the process of finding the EQUSFs of a FFF field. The

scatter functions of a 10 MV WFF, a 10 MV FFF and a 10 MV

VUFFF beam are plotted as a function of field size. In this

example the 10 x 10 cm² reference field of the 10 MV FFF

beam has the same scatter contribution as the 9 x 9 cm²

field of the 10 MV VUFFF beam which was defined as

EQUSFs. The scatter function of the 10 MV WFF was

plotted for comparison purposes.

Conclusion

It has been shown that with the introduction of a VUFFF

beam, EQUSFs can be consistently defined for a variety of

energies and collimations. These EQUSFs serve as basis for

a unified reference dosimetry protocol for all different

types of FFF machines.

References

[1] Fogliata A, Fleckenstein J, Schneider F, et al.

Flattening filter free beams from TrueBeam and Versa HD

units: Evaluation of the parameters for quality assurance.

Med. Phys. 2016;43.

[2] Chang Z, Wu Q, Adamson J, et al. Commissioning and

dosimetric characteristics of TrueBeam system:

Composite data of three TrueBeam machines. Med Phys.

2012;39:6981–7018.

OC-0531 The influence of detector resolution on pre-

treatment quality assurance in SBRT

A. Bruschi

1

, S. Russo

2

, M. Esposito

2

, S. Pini

2

, A. Ghirelli

2

,

G. Zatelli

2

, P. Bastiani

3

1

University od Florence, University of Florence, FIrenze,

Italy

2

Azienda USL Toscana Centro- Sede di Firenze, Medical

Physics Unit, Florence, Italy

3

Azienda USL Toscana Centro- Sede di Firenze,

Radiotherapy Unit, Florence, Italy

Purpose or Objective

2D detector arrays have become the standard device for

verification of VMAT dose distributions. The detector pixel

size is a key parameter to reproduce complex dose

distributions.

Aim of this work is to compare the gamma analysis of three

ionization chamber systems and to test the ability of each

system to detect deliberate errors.

Material and Methods

Measurements performed by PTW Octavius 4D 729

(5x5x5mm

3

ionization chamber, 10 mm spacing), PTW

Octavius 4D 1500 (4.4x4.4x3mm

3

ionization chambers, 7.1

mm spacing) and PTW Octavius 4D 1000 SRS (2.3x 2.3x 0.5

mm

3

liquid filled ionization chamber, 2.5 mm spacing) in

the PTW Octavius 4D phantom were used to validate the

dosimetric accuracy of the VMAT delivery. Firstly, 50

VMAT SBRT treatment plans from a variety of clinical sites

were considered. Secondly, systematic variations in

collimator (2° and 5° rotation) and gantry angle (shift of

2° and 5°) and lack of monitor units were applied to four

clinical treatments (2 lung tumors, 1 spine and 1

abdominal lymph node) in order to establish the detection

sensitivity of the three devices. Measurements were

compared with TPS Elekta Monaco computed doses via

local gamma analysis ( 2%L/ 2 mm, 2%L/ 1 mm and 1% L/1

mm). For the 729 and 1500 detectors, the resolution was

improved by merging two measurements performed with

5 mm couch shift.

The threshold for a success in error detectability was

established, by using the concept of confidence limit (CL),

as suggested by AAPM Task Group 119 [1]:

CL = (100 - D) + σ

where D and σ are respectively the mean dose and the

standard deviation of the distribution of the gamma

passing rate (35 plans for 6 MV and 15 plans for 10 MV)

measured by each dosimeter; the detectability threshold

(DT) has been calculated as:

DT = 100 - CL

Results

The average pass rate with 2%L/2 mm criterion for the 6MV

cases was 86.6 ± 5.2 (no shift) and 90.3 ± 4.3 (merged) for

729, 91.5 ± 3.7 (no shift) and 94.7 ± 2.9 (merged) for 1500

and 98.9 ± 1.1 for 1000 SRS. Box plot relative to 6 MV VMAT

SBRT 2%L/2 mm pass rate is presented in figure 1. Similar

results were achieved for 10 MV plans. The results for the

plans with errors, normalized to the DT for a success in

error detectability, show that only the SRS system can

distinguish the delivery errors, as shown in figure 2 for

6MV. Only a 2%L/1 mm gamma criterion proved to be

sensitive enough to detect errors.

Figure 1 : Box plot for 2%L/2 mm gamma passing rate of 6

MV VMAT SBRT plans for the investigated 2D arrays.