Table of Contents Table of Contents
Previous Page  438 / 1082 Next Page
Information
Show Menu
Previous Page 438 / 1082 Next Page
Page Background

S423

ESTRO 36 2017

_______________________________________________________________________________________________

Gesellschaft für Schwerionenforschung (GSI) were used as

reference data.

Results

For all physics lists, the relative dose differences up to the

Bragg peak were found to be less than 4% compared to

measurements. Beyond the Bragg peak, in the so-called

fragmentation tail, differences increased notably, by up

to one order of magnitude. However, the absolute dose

difference in the fragmentation tail was comparable to

the absolute difference before the Bragg peak. The QMD

model systematically overestimated whereas the other

models underestimated the dose in the fragmentation tail.

Overall, deviations to the measurement were less than 2%

of the maximum dose for all models, disregarding the dose

fall off region due to the steep dose gradient.

Partial charge changing cross sections simulated with the

BIC, BERT and QBBC models deviated up to 60% from the

measurements, INCLXX up to 38% and the QMD model up

to 24%. However, the significance on fragmentation in

particle therapy is limited by the high energy equal to 630

MeV/u used in the measurements.

Conclusion

IDDs simulated with Gate/Geant4 agreed well with

measurements for all models under investigation,

although notable deviations were observed in the

fragmentation tail. Measured partial charge changing

cross sections could best be reproduced using the QMD

model, whereas the BIC model showed considerable

discrepancies. Therefore, Gate/Geant4 can be considered

a valid dose calculation tool for oxygen ion beams and will

further on be used for the development of a pencil beam

algorithm for oxygen ions. The QMD model is

recommended in order to obtain accurate fragmentation

results, which is essential for radiation oncology purposes.

PO-0802 Experimental validation of single detector

proton radiography with scanning beams

C. Chirvase

1

, K. Teo

2

, R. Barlow

1

, E.H. Bentefour

3

1

International Institute for Accelerator Applications, The

University of Huddersfield, Huddersfield, United

Kingdom

2

University of Pennsylvania, Department of Radiation

Oncology, Philadelphia PA, USA

3

Advanced Technology Group, Ion Beam Applications

s.a., Louvain-la-Neuve, Belgium

Purpose or Objective

Proton radiography represents a potential solution to solve

the uncertainties of dose delivery in proton therapy. It can

be used for in-vivo beam range verification; patient

specific Hounsfield unit (HU) to relative stopping power

calibration and improving patient set-up. The purpose of

this study is to experimentally validate the concept of the

energy resolved dose measurement for proton radiography

using a single detector with Pencil Beam Scanning (PBS).

Material and Methods

A 45 layers imaging field with a size of 30 x 30 cm

2

and

energies between 226 MeV and 115 MeV is used to deliver

a uniform dose. The dose per spot is 4.25 mGy with spot

spacing equal to the beam sigma. The imaging field is first

delivered on wedge shaped water phantom to produce

calibration library of Energy Resolved Dose Functions

(ERDF) between 0 cm and 30 cm. Then, the same imaging

field is delivered in three different configurations: a stack

of solid water in a stairs shape with thicknesses between

1 mm and 10 mm – that determines the accuracy with

which the WEPL (water-equivalent path length) can be

retrieved, CIRS lung phantom – that illustrates the

accuracy on the density of multiple materials and a head

phantom – which represents a realistic case of

heterogeneous target.

As shown in Figure 1, proton radiographs are recorded with

a commercial 2D detector (Lynx, IBA-Dosimetry,

Schwarzenbruck, Germany) which has an active area of

300 × 300 mm² with an effective resolution of 0.5 mm.

Figure 1. Experimental set-up with an example of proton

radiograph imaged at beam energy of 220 MeV.

Results

In this study we demonstrate the robustness of the energy

resolved dose measurement method for single detector

proton imaging. It shows the capability to determine the

WEPL with sub-millimeter accuracy in a homogeneous

target and performs well in heterogeneous target, proving

an accuracy better than 2 mm even in most heterogeneous

areas of a head phantom. These performances are

achieved by using an imaging field with as little as 5

energy layers with spacing up to 10 mm between the

layers.

Although the optimization of the imaging dose was not a

goal of this study, only ~21 mGy per cm

2

is sufficient to

obtain the above accuracies. This dose can be further

decreased by using a detector with higher sensitivity and

by reducing the number of beam spots per layer of the

imaging field.

Conclusion

Proton radiography with single detector using energy

resolved dose measurement did show potential for clinical

use. Further studies are needed to optimize the imaging

dose and the clinical workflow.

PO-0803 CloudMC, a Cloud Computing application for

fast Monte Carlo treatment verification

H. Miras

1

, R. Jiménez

2

, R. Arrans

1

, A. Perales

3

, M. Cortés-

Giraldo

3

, A. Ortiz

1

, J. Macías

1

1

Hospital Universitario Virgen Macarena, Medical Physics,

Sevilla, Spain

2

Icinetic TIC SL, R&D division, Sevilla, Spain

3

Universidad de Sevilla, Atomic- Molecular and Nuclear

Physics Department, Sevilla, Spain

Purpose or Objective

CloudMC is a cloud-based solution developed for r educing

time of Monte Carlo (MC) simulation s through

parallelization in multiple virtual computing nodes in the

Microsoft’s cloud. This work presents an update for

performing MC calculation of complete RT treatments in

an easy, fast and cheap way.

Material and Methods

The application CloudMC, presented in previous works, has

been updated with a solution for automatically perform

MC treatment verification. CloudMC architecture (figure

1) is divided into two units. The processing unit consists of

a web role that hosts the user interface and is responsible

of provisioning the computing worker roles pool, where

the tasks are distributed and executed, and a reducer

worker role that merges the outputs. The storage unit

contains the user files, a data base with the users and

simulations metadata and a system of message queues to

maintain asynchronous communication between the front-

end and the back-end of the application.