S426
ESTRO 36 2017
_______________________________________________________________________________________________
dosimetric end-to-end procedures for protons based on
customized anthropomorphic phantoms and different
dosimetric techniques.
Material and Methods
A homogeneous polystyrene phantom and two
anthropomorphic phantoms (pelvis and head phantom)
have been customized to allocate different detectors such
as radiochromic films, ionization chambers and alanine
pellets. During testing, the phantoms were moving
through the workflow as real patients to simulate the
entire clinical procedure. The CT scans were acquired with
pre-defined scan protocols used at MA for cranial and
pelvic treatments. All treatment planning steps were
performed with RayStation v5.0.2 treatment planning
system (TPS). A physical dose of 10 Gy was planned to
clinically shaped target volumes in order to achieve
uniformity better than 0.5% on the dose delivered to the
alanine pellets. In the treatment room the plans were
delivered to the phantoms loaded either with alanine
pellets and radiochromic EBT3 films (figure 1) or two
Farmer chambers. The alanine pellets (5.0 mm diameter
and 2.3 mm thickness) and their read-out were provided
by the National Physical Laboratory (NPL). One of the
challenges of alanine for dosimetry in particle beams is
the known response dependency (quenching) on the
charge, the fluence and the energy of the particles
constituting the mixed radiation field. Corrections for this
were derived by a Monte Carlo dose calculation platform
implemented in a non-clinical version of RayStation.
Results
The measured absolute dose to water obtained with the
Farmer chamber in all delivered plans was within 2% of the
TPS calculated dose. A lateral 2D homogeneity of 3% inside
the treatment field was measured with EBT films. Doses
determined with the alanine pellets after correction for
the quenching effect showed a mean deviation within 3%
and a maximum deviation below 7% in the homogeneous
and anthropomorphic phantoms.
Conclusion
The end-to-end test procedures developed at MedAustron
showed that the entire chain of radiation treatment works
efficiently and with accurate dosimetric results. Our
experience shows that alanine pellets are suitable
detectors for dosimetry audits and developed procedures
can be used to support implementation of scanning beam
delivery technology in clinical practice
.
PO-0807 Practical advantages of a transmission
chamber in relative dosimetry of Brainlab conical
applicators
B. Tang
1
, J. Li
1
, S. Kang
1
, P. WANG
1
, L.C. Orlandini
1
1
Sichuan Cancer Hospital, Radiation Oncology, Chengdu,
China
Purpose or Objective
The commissioning of a radiosurgery unit requires the
acquisition of specific detectors able to characterize the
geometry and dosimetry of small fields. The acquisition of
the equipment for absolute dosimetry remains the priority
for the Hospitals, considering that relative measurements
can be performed without a reference chamber using long
acquisition time. The commissioning results therefore in a
tedious procedure. In this study, a new transmission
chamber (Stealth Chamber, IBA Dosimetry) was used as a
reference chamber (RC) in relative dosimetry of Brainlab
cone applicators. The timing of the practical procedure
and dosimetry results with and without the reference
chamber, will be analyzed and compared.
Material and Methods
IBA SFD3G diode detector was used to measure the 6MV
photon beam of a Varian Novalis used with Brainlab cone
applicators. Inline and crossline profiles at different
depths and central axis depths doses (PDDs) were
measured with a motorized water phantom (Blue
phantom, IBA Dosimetry) and OmniPro v7.4 software for
every cone. The measurements were acquired with the
transmission reference chamber positioned on the gantry
head in a continuous mode and without RC in a step by
step mode. The details of the acquisition parameters were
reported in Table 1. The total measurements time for each
procedure was registered.
Table 1 The Acquistion parameters with and without
stealth chamber
Acquisition
parameters/method
No
reference
Stealth
Chamber
Scan mode
step by step continuous
Scan speed
-
5 mm/s
In-scan positioning speed 5 mm/s
-
Positioning speed
10 mm/s
10 mm/s
Acquisition time
5 s
-
Stabilzation time
1 s
0.08 s
Results
Profiles at depth 10 cm for 4/15 mm diameter cones and
the depth doses acquired with the two procedures (Figure
1&Figure 2) shown a good agreement. The total
measurement time registered was 490 seconds for the
PDDs acquisition without RC and only 64 seconds when the
scan mode change from “step by step” to “continuous”
after stealth chamber was in place. The overall
measurement time for 4mm diameter was 575 s and 12 s
without and with RC respectively, 735 s and 17 s for the
15 mm diameter cone.
Figure 1
Figure 2