Table of Contents Table of Contents
Previous Page  93 / 1082 Next Page
Information
Show Menu
Previous Page 93 / 1082 Next Page
Page Background

S80

ESTRO 36 2017

_______________________________________________________________________________________________

presence of metal artefacts. To eliminate left-right bias,

each combination was shown twice.

Results

VMDE images reconstructed at energies in the range 60 to

70 keV showed improved CNR for all soft tissue regions

when compared to the standard CBCT. On average, the

reconstruction energy corresponding to the maximum CNR

improvement is 65.5 ± 2.4 keV. The increase in maximum

CNR varied from 29% to 78%.

The clinical observer comparison gave a series of rankings

for each image series for each patient (see table 1). Using

signed rank Wilcoxon comparison test, the observers found

the VMDE images at 65 keV preferable to the standard

CBCT image. The p-value was found to be < 0.01, where p

< 0.05 is considered significant. An estimate of inter-

observer variability test was done with Fleiss’ kappa and

found to be moderate with a κ-value of 0.47, where values

above 0.4 is considered acceptable and 1 is perfect

agreement. Occasionally, an observer ranked the 75 keV

reconstruction as the most preferable image while the

overall preferred image was the 65 keV reconstruction.

Except in the case of patient one where the standard CBCT

image was ranked the highest of all.

Conclusion

VMDE images can increase soft tissue contrast and improve

clinical image quality for image-guided radiotherapy

compared to the standard CBCT protocol.

OC-0160 Radiomics Features Harmonization for CT

and CBCT in Rectal Cancer

R. Luo

1

, J. Wang

1

, H. Zhong

1

, J. Gan

1

, P. Hu

1

, L. Shen

1

,

W. Hu

1

, Z. Zhang

1

1

Fudan University Cancer Hospital, Radiation Oncology,

Shanghai, China

Purpose or Objective

Inter-scanner variability can lead to degradation of

radiomics

model

quality.

Therefore,

feature

harmonization is necessary for consistent findings in

radiomics studies, especially for multi-institution studies.

The purpose of this study is

to establish harmonization

methods for CT and CBCT radiomics features in rectal

cancer, and to provide a harmonization evaluation

method.

Material and Methods

Three harmonization strategies were tested in this study,

including no correction, simple correction and phantom

based correction. 50 rectal cancer patients with both

planning CT images and positioning CBCT images before

the first fraction of treatment were collected for

harmonization performance evaluation. 203 features were

extracted from CT and CBCT images. For the phantom

based correction, a texture phantom comprised of 30

different materials was designed for features selection

and nonlinear functions generation for normalizing CT and

CBCT features.The Main workflow was shown in Figure 1.

Mixed datasets consisting of CT and CBCT features were

generated for harmonization performance evaluation

using cluster analysis. The harmonization performance

was evaluated by Chi-square testing between clustering

results and scanner machines, and the clustering

consistency with original CT feature. These tests were

repeated for 50 times with randomized sample

generation.

Figure 1. Main Workflow. Four steps of this study:(I)

Feature selection by features range comparison.

(II)Feature selection by spearman correlation test. (III)

Nonlinear mapping function generation using texture

phantom. (IV)Correction methods performance evaluation

on patients.

Results

41 of the 203 radiomics features were selected by range

comparison and spearman correlation test. Among 50

randomized sampling processes, all clustering (100%)

results without any correction showed high correlation

with imaging machine (p>0.05, χ^

2

test), while this

probability reduced to 0 % and 42% respectively when

simple correction or phantom based correction were

applied. Average accuracy and Kappa index increased

significantly (p<0.05, t-test), respectively to 0.71±0.07

and 0.42±0.12 for simple correction method and 0.68±0.06

and 0.36±0.14 for phantom based correction method, from

0.61±0.06 and 0.23±0.13 without any correction.

Table1. Performance evaluation result for different

harmonization strategies.

Conclusion

This is the first study focused on feature harmonization for

CT images. Two proposed correction methods, simple

correction and phantom based correction, were verified

to be feasible for CT and CBCT harmonization, which could

significantly improve the modeling consistency.

Proffered Papers: Novelties in image guidance

OC-0161 patient tolerance of stereotactic MR-guided

adaptive radiation therapy: an assessment using PRO’s