Table of Contents Table of Contents
Previous Page  957 / 1082 Next Page
Information
Show Menu
Previous Page 957 / 1082 Next Page
Page Background

S941

ESTRO 36 2017

_______________________________________________________________________________________________

An optimization bolus was added from the body and 12 mm

expansion, defined as mass density 1.0 g/cm

3

. The

prescribed dose was 12 Gy delivered in 6 fractions. The

dose planning was aimed to keep D

95%

> 95% to PTV and

minimizing dose to organs at risk, which was defined as

the rest of the body. An optimization structure was used

to create a tangential irradiation of the skin, minimizing

the dose to normal tissue. We tested the bolus effect of

Neoprene with Gafchromic EBT3 film by irradiating slabs

of 7mm, dry and soaked in water. To verify skin doses, the

phantom with wet suit was irradiated with several 2x2 cm

2

slabs of film taped to the body. The film were evaluated

at least 24 hours after irradiation. Corresponding detector

array measurements (Delta4, Scandidos) were done and

evaluated with gamma analysis. Further, a robustness test

was done by moving the phantom 10 mm in the x, y and z

directions, to evaluate the effect of mispositioning.

Results

Results of planning and robustness tests are presented in

table 1. Measured data fit to depth dose data yields a dose

maximum at 28 mm for Neoprene. Hence, 7 mm is

equivalent to 3 mm thick water bolus and lightly soaked

Neoprene adds another 1.2 mm thickness of water. Delta4

gamma analysis with 2 mm and 3%, global dose, is clinical

acceptable with regards to deliverability (M = 93%, SD =

3%). The verification of 27 film slabs for skin dose gave an

average difference from TPS dose of 4% (SD = 3%), figure

1.

Conclusion

The difference of measured dose compared to TPS, for

both film and Delta4 dosimetry is larger than most types

of targets treated with HT which is to be expected

considering the technique and size of target. The

deliverability is within limit of our clinic action levels

(gamma pass rate < 90%) and neoprene is feasible as bolus.

The benefits, in comparison with reported electron

treatments, are target homogeneity and target coverage

with good immobilization and complete irradiation with

two positions. The higher dose to organs at risk than

reported with electrons needs to be addressed if

acceptable with regards to

toxicity.

EP-1736 Radiation and lasers isocenters coincidence

with ArcCheck phantom

F. Tato de las Cuevas

1

, J. Yuste Lopez

1

1

Hosp. Univ. de Canarias, Medical Physics Dept., Santa

Cruz de Tenerife, Spain

Purpose or Objective

One tool of Machine QA module of ArcCheck phantom (

AC

)

software checks Radiation and Lasers Isocenters

Coincidence (

RLIC

). The purpose of this work is to

evaluate the precision and accuracy of this software tool,

comparing it to the same test made with EPID (Electronic

Portal Imaging Device).

Material and Methods

The LINAC is an Elekta Synergy with Agility MLC and 6 MV

energy.

The RLIC with ArcCheck phantom (

AC

) are obtained

following the instructions of the software manual. The

measurements are done in continuous gantry movement

and for discrete gantry angles. Measurements are made at

9 º collimator angle for a 1x25 cm field. A series of

measurements were made also in 99º to see the MLC

effect, as Agility head has not backup jaws. The AC

displacements from laser isocenter in two directions are

made in order to check software sensitivity.

RLIC are made with EPID, positioning a Bearing Ball (

BB

)

in the lasers isocenter of a 5x5 cm field and acquiring

Images from 0º to 360º gantry angles in 45 º steps. The

radiation center of the squared field and the center of the

BB are calculated with a MATLAB in-house software. BB

center is calculated with sub-pixel accuracy in each

direction, 3 profiles are obtained and fitted to Gaussian

curves, and the mean maximum of the 3 curves is

calculated. Radiation field center is obtained calculating

the 50% pixel value of a vertical and horizontal profile.

The difference between BB center and radiation field

center are computed for each gantry angle for in-plane

and cross-plane directions. The RLIC for EPID

measurements are computed using these values.

Results

The RLIC results obtained with AC for each gantry are

compared with EPID in the first figure. The mean distance

over all gantry angles, for AC (for 9 and 99 º collimator

degrees) and EPID are: 0.3, 0.6 and, 0.7 mm, respectively.

The AC results are just distance (because this phantom is

not capable of give deviation in in-plane direction for each

gantry angle). The results for AC for 9º are higher than for

99º because of the irregular MLC radiation field limit

exposed for 9º to the AC diodes. The RLIC for EPID are

given in in-plane and cross-plane directions, the distance

for each gantry angle is calculated from both directions

and show a bigger mean value than for AC, because of

being calculated in just one direction in this phantom.