S942
ESTRO 36 2017
_______________________________________________________________________________________________
Table 1 shows RLIC results for AC and EPID. AC data is
given by phantom software. It can be noticed that the
coincidence for both isocenters is lower for the EPID, this
can be explained because AC just take into account one
direction in each gantry angle
.
The sensitivity of AC for RLIC is fairly good taking into
account the uncertainties of measurement 1 mm between
laser positions.
Conclusion
ArcCheck software is capable of give a fairly accurate
measurement of the laser and radiation isocenters
coincidence, taking into account to add about 0.5 mm
displacement in X and Z directions.
EP-1737 Efficient troubleshooting of accelerator faults
using the TrueBeam Log Viewer software application
L.H. Praestegaard
1
1
Aarhus University Hospital, Department of Oncology,
Aarhus, Denmark
Purpose or Objective
In case of an accelerator fault, the identification of the
root cause often takes much longer time than the
correction of the fault itself (for example replacement of
a component or a calibration). Accordingly, the uptime of
an accelerator very much depends on an efficient
troubleshooting process. In addition, an overview of
existing faults is essential for an efficient planning of
service tasks.
Material and Methods
The TrueBeam Log Viewer application is an in-house
software application developed in C#. For the Varian
TrueBeam system all faults occurring during beam on
produces an event. With the application all events, for
user-selected accelerators and data interval, can be listed
including fault details, treatment plan details, mechanical
axes, imaging parameters, and imaging arm positions at
the time of the event (see figure 1). For each event the
TrueBeam system produces a number of node records,
each with detailed information about system parameters
versus time just before the fault occurred (for example all
MLC positions and motor currents at 500 time steps of 10
ms). All node records are readily available in the
application from the list of events. In addition, the
application can generate an event alarm, including the
corresponding event data, each time an event occurs for
user-selected accelerators.
For each dynamic beam delivery the TrueBeam system
also produces a single log of the main system parameters
in a trajectory log text file. The TrueBeam Log Viewer
application can list these parameters versus time for user-
selected accelerators and data interval.
All lists can be sorted with respect to any column, non-
clinical events can be ignored, events not matching a user-
selected main fault code or fault description string can be
ignored, and all lists can be exported to Excel for further
analysis.
Results
Sorting the list of events with respect to the main fault
code provides a fast overview of the number faults for
each fault type and accelerator, facilitating an easy
prioritization of faults (see figure 1). In addition,
treatment plan details and mechanical axes for each event
(for example beam energy, the number of monitor units,
the patient ID, and the gantry angle) are valuable for the
identification of the root cause of the corresponding fault.
A more detailed insight into the root cause of a fault can
be gained by analyzing node records of the corresponding
event in Excel. An example of a node record for a gantry
fault is shown in figure 2. Also trending of the main system
parameters from the trajectory log text files with Excel is
a strong troubleshooting tool.
Conclusion
The TrueBeam Log Viewer application is an efficient tool
both for obtaining an overview of existing faults and for
identification of the root cause of faults. In our clinic the
application has provided a much more substantiated
prioritization of service tasks and faster identification of
the root cause of faults. Varian has no service tools with
this capability.
EP-1738 Performance of a new EPID panel and
opportunities for a fast MV-CBCT acquisition.
C. Kennedy
1
, C. Ling
2
, R. Scheuermann
1
, D. Mihailidis
1
, J.
Metz
1
1
University of Pennsylvania, Radiation Oncology,
Philadelphia, USA
2
Varian Medical Systems, Reserach and Development,
Palo Alto- CA, USA
Purpose or Objective