Table of Contents Table of Contents
Previous Page  971 / 1082 Next Page
Information
Show Menu
Previous Page 971 / 1082 Next Page
Page Background

S955

ESTRO 36 2017

_______________________________________________________________________________________________

quality have been assessed. Subsequent real-time

monitoring issues have been considered.

Results

The presence of IQM chamber in the beam path changed

the pass rate of MapCheck (IMRT), and ArcCheck (VMAT)

within ±1% when %Dose–DTA criteria were employed, as

shown in Table-1. The calculated and measured IQM

signals on a Varian TrueBeam Unit for 340 randomly chosen

IMRT field segments from clinical plans show good

agreements (Fig.1): 95% of segments are within ±3; total

cumulative signals for VMAT delivery were within± 2%.

By introducing the system as a pre-treatment QA tool 700

hours of staff time and 180 hours of machine time can be

saved annually, for a facility treating 240 IMRT and VMAT

patients per day. Additionally, the segment by segment

dosimetry and independent gantry angle monitoring

provides added quality values for pre-treatment QA.

Daily monitoring of beam delivery may save more machine

and staff time by eliminating pre-treatment QA, while

improving patient safety. However, a number of issues

need to be addressed, such as: (1) Modification of TPS

beam model to include the effect of the IQM chamber on

the beam (2) The TPS should allow an accessory code in

IMRT and VMAT (3) The Linac manufacturer should make

an accessory code available for the on-line monitor.

Conclusion

Clinical implementation of the system in multiple phases

helps understanding the performance characteristics of

the system, allows smooth transition of QA practices,

make overall clinical workflow safe and effective for real-

time beam monitoring.

EP-1759 MLC positioning study based on EPID images

analyzed with the Dosimetry Check software

C. Avigo

1

, M. Mignogna

2

, S. Linslata

3

1

National Research Council, Institute of Clinical

Physiology, Pisa, Italy

2

Azienda USL Toscana nord ovest- S. Luca Hospital,

Radioterapia, Lucca, Italy

3

Azienda USL Toscana nord ovest- S. Luca Hospital, Fisica

Sanitaria, Lucca, Italy

Purpose or Objective

The IMRT requires extensive knowledge of the MLC

position accuracy and repeatability since when accurate

leaf positioning is lost significant dose delivery errors can

occur. Therefore, the MLC QA is crucial for a complete

control of the patient treatment. The use of EPID for this

scope can be very helpful in saving time providing images

with high spatial resolution and directly digitalized.

Dosimetry Check is a commercial software which uses EPID

images for pre-treatment verification, in vivo-dosimetry

and also MLC QA. The aim of this work was to validate the

combined system EPID-Dosimetry Check, in order to

control the leaf positioning of an Elekta Agility MLC.

Material and Methods

The leaf position is defined as the position of the 50% of

the dose profile. This measurement depends on relative

position of the beam source and of the leaves. In order to

validate the EPID measurements of the absolute leaf

positions, 10 dose profiles, at the center of different leaf

pairs of the same MLC field, were acquired with an Elekta

iViewGT EPID and with a diode positioned in a water

phantom. The comparison between the two detectors was

performed by Matlab. Garden Fence (GF) was chosen as

test of the leaf position accuracy and a preliminary study

on the gap width was conducted. Leaf position accuracy

was checked automatically with DC by acquiring GF at the

4 cardinal gantry angles and with all the beam energies (6,

10 and 15MV), while the reproducibility was tested with 5

GF repeated in one day and 6 repeated in a time interval

of 70 days.

Results

The difference between EPID and diode absolute

measurement of the leaf positions was less than 0.8mm

for all the analyzed leaves, resulting from the summation

of an error due to the isocenter identification (0,5mm)

plus the leaf positioning error (0.2mm). The gap width

study revealed that, because of the penumbra widening

observed in small fields, the leaf position could be

accurately measured as the 50% of the edge profile, only

if the gap width is equal or larger than 16mm with 6MV

beam. Therefore, GF with 20mm gap was chosen as leaf

position accuracy test for all the energies in order to

distinguish the effect of beam source from that of leaf

positioning. For the GF at different gantry angles the

difference between the measured and the prescribed

position was well within 1.0mm for all the leaves.

Moreover, reproducibility of each leaf position resulted to

differ from its average value less than 0.4mm.

Conclusion

This work permitted to assess the accuracy and the

repeatability of the Elekta Agility MLC leaf positioning by

the combined use of the Elekta IviewGT EPID and the

Dosimetry Check software through the acquisition and the

analysis of Garden Fence test. This system was validated

comparing the EPID with a diode in a water phantom and

assessing the minimum gap width necessary for an

accurate leaf position measurement at all energies which

is useful to distinguish issues related to beam symmetry

from those related to leaf positioning.

EP-1760 A simple method for estimating the

longitudinal isocentre shift due to gantry motion

R. Hudej

1

, D. Brojan

1

, S. Pulko

2

, P. Peterlin

1

1

Institute of Oncology Ljubljana, Department of

Radiophysics, Ljubljana, Slovenia

2

University Clinical Centre Maribor, Department of

Oncology- Radiotherapy Unit, Maribor, Slovenia

Purpose or Objective

The isocentre as a point of intersection of the three

rotational axes (gantry, collimator and treatment couch)

ideally remains fixed in space during the rotation of

gantry, collimator, or the treatment couch. Due to the

mechanical limitations, gantry sags slightly, and

consequently the radiation isocentre shifts slightly

towards the treatment couch when the gantry rotates

from the uppermost to the lowermost position. The

purpose of this study is to assess this shift.

Material and Methods

A strip of radiochromic film embedded in a suitable water-

equivalent phantom is irradiated with a cross-line half-slit

field from the top (0°). Then the gantry is rotated to the

lowermost position (180°) without moving the jaws and

the phantom is irradiated again. The film is scanned and

analysed with an image analysis script. The central lines

of both half-slit images are determined, then the

intersection angle between them is calculated, and finally

the distance between the intersections of extrapolated

lines with the 'sagittal” plane is calculated.