Independent Laboratory Validation

The independent validation verifies the method performance observed in 

the method developer’s laboratory using viable Bacillus anthracis cultures.  

The independent laboratory assessed  method performance including 

1) probability of detection (POD) at the acceptable minimum detection level (AMDL) 

or lower

2) inclusivity

3) exclusivity

4) environmental interference  

Certified reference materials were used for POD repeatability dynamic range and specificity

testing. Other reference materials are used for inclusivity, exclusivity and environmental 

factors testing.

Robustness testing (instruments and people)

Shelf life testing of Developer Kits

Multicenter Validation Testing

Identify test centers

A

li

i h CDC l i

d F d l l

ssure comp ance w t 

 regu at ons an e era aw

Identify individuals not associated with developer or independent 

laboratory to do the testing. 

Determine the acceptable skill of testers

Determine when testing will be done

Assure administrators, safety officers and support staff are qualified, aware and 

support the project

Have a well vetted test plan in place with easy data recording instructions

Assure samples can be sent to receiving laboratories

level Randomize and blind code the subsamples

Conduct testing over a short period of time if possible

Record results (pos or neg) for each primer pair and each sample