Table of Contents Table of Contents
Previous Page  9 / 35 Next Page
Information
Show Menu
Previous Page 9 / 35 Next Page
Page Background

1352 

Lacorn & Weiss

: J

ournal of

AOAC I

nternational

Vol. 98, No. 5, 2015

J. Criteria for Acceptance of the Standard Curve

The shape of the standard curve is shown in the quality

assurance certificate enclosed in the test kit. Absorbances may

vary between different runs (e.g., due to different temperatures

or analysts). However, the shape of the standard curve should

be similar to the one given in the quality assurance certificate.

Minimum requirements are as follows:

(

1

) OD at 450 nm for standard 1 higher than 0.8.

(

2

) OD values for standards should continuously decrease

with higher concentrations, especially when comparing

standard 1 (0 ng/mL) and standard 2 (20 ng/mL).

(

3

) An OD value for standard 1 that is much higher than

the OD value stated in the certificate could be an indication of

errors during pipetting or incubation.

Results and Discussion

Collaborative Study Results

After finishing the analysis, each participant sent the data to

the Study Coordinator. These results are given in Table 1. After

statistical analysis of the data set, three problem laboratories

were identified. Further review found Laboratory F did not

run the calibrators in duplicate determinations as directed.

Laboratory E found no difference between calibration standards

S1 and S2, and as a consequence, a high OD difference

between standards S4 and S5 led to an unusual curve shape. An

interview with Laboratory E also revealed technical problems

during sample preparation. Laboratory K had a variation in the

calibration curve that was too high, and an interview revealed

the possibility of gluten contamination in the laboratory and

incorrect pipetting. As a result of these deviations, all data from

Laboratories E, F, and K were excluded from the statistical

evaluation.

For sample 5 (naturally contaminated syrup), all values were

calculated by cubic spline. Due to the fact that some OD values

were below the OD values of standard 2 (10 ng/mL prolamin;

corresponds to concentration of 10 mg/kg in the sample), these

values were extrapolated by the software. For the gluten-free

samples 1 and 4 the RIDA

®

SOFT Win software returned only

a result of <10 mg/kg, and extrapolation led to unrealistic

values. To be able to use the results of the analysis of the gluten-

free samples 1 and 4 in the performance statistics, estimates

of concentration values for these samples were required. For

this purpose, the calibration curves were constructed by using

a second-order polynomial model and used to recalculate the

results for samples 1 and 4 (7). This calibration provided an

estimate of concentrations for the gluten-free samples (Tables 1

and 2).

Statistical Analysis and Discussion

The remaining data of 13 laboratories are shown in Table 2

and were used to calculate the necessary statistics. Only

three outlying values were identified according to AOAC

INTERNATIONAL guidelines (12). These are indicated in

Table 2 by the superscripts “c” (for a Cochran outlier) and

“d” (for a double Grubbs’ outlier). The performance statistics

without outliers are shown in Table 2015.05.

From the measured overall mean concentrations of the

gluten-containing samples, recovery rates were calculated.

Table 2. Gluten concentrations determined by R5 competitive ELISA after eliminating laboratories E, F, and K

Gluten concentration, mg/kg

a

1

b

2

3

4

5

6

7

Repeat

Lab

1

2

1

2

1

2

1

2

1

2

1

2

1

2

A

2.13 5.80

23.6 20.5

111.6 93.9

4.47 7.73

7.6 8.62

46.7 47.2

153.0 170.0

B

1.46 2.66

40.8 13.8 151.4 127.4 2.98 2.13

10.6 5.1

38.8 53.0

163.6 122.8

C

5.30 10.6

34.2

c

82.2

c

192.2 107.6 6.12 1.90

12.8 12.6

47.2 67.4

181.4 143.4

D

0.74 1.77

23.8 28.6 175.2 97.6 –3.35 –3.41

9.8 11.0

33.0 60.2

106.4 107.6

G

6.06 4.30

32.4 32.0 216.2

d

208.2

d

3.29 –2.34 15.0 14.0

46.8 85.4

192.8 203.0

H

7.02 1.56

44.4 26.2 145.6 32.8

5.79 3.17

20.5 16.1

38.8 31.1

94.6 88.9

I

–0.65 –1.33 22.2 13.8 101.2 64.4 –0.89 –0.62

5.4 4.2

35.8 45.0

118.4 75.0

J

–1.50 1.14

21.2 20.0 121.8 128.8 –0.73 –1.63

7.4 8.0

45.6 58.3

132.9 139.2

L

1.69 –0.33 39.8 49.0 224.8

d

228.8

d

–1.83 3.39

13.2 11.6

64.0 67.2

171.6 244.6

M –0.66 4.13

19.9 19.3 129.4 133.6 –2.27 –0.62 10.0 8.6

36.1 39.6

161.7 120.4

N

0.04 0.76

34.2 18.4

97.0 108.6 1.84 4.41

10.8 9.2

43.4 44.6

117.6 154.4

O

1.57 0.41

19.1 16.5

110.7 136.6 –0.11 1.26

11.7 8.2

51.3 46.3

152.8 164.6

P

5.96 0.84

25.4 24.8 149.4 111.2

1.54 1.33

12.6 10.8

38.2 46.2 194.8 111.2

a

 The calculation of the concentrations of the gluten-containing samples 2, 3, 5, 6, and 7 was done on the basis of a cubic spline function using the

RIDA

®

SOFT Win software; the statistics of the gluten-free samples 1 and 4 were calculated on the basis of a second-order polynomial function;

values for blinded samples are given as repeat 1 or repeat 2.

b

 For samples 1–7

see

Table 1.

c

 Means outlier according to the Cochran test.

d

 Means outlier according to the double Grubbs’ test.