Table of Contents Table of Contents
Previous Page  570 / 670 Next Page
Information
Show Menu
Previous Page 570 / 670 Next Page
Page Background

Page

8

of

39

breakthrough of borate to the analytical column degrading method performance (primarily observed in the peak

asymmetry (EP) of the glucoheptose (also known as the USP tailing factor).

2.

Injection

A single 4 μL injection of each sample test solution is made. At the start of each sequence, 6 precision

injections are made to ensure that system is stable, followed by the 4 levels of working standard. Samples are

bracketed with working standards after every 13 injections. Autosampler is maintained at 10 °C.

3.

Electrochemical detector parameters

This method utilizes the carbohydrate triple waveform referred to in

Thermo/Dionex Tech Note 21, Waveform B (4).

Note this waveform is not appropriate for disposable gold

electrodes!

The reference electrode is set to AgCl mode.

4.

Retention times

Typically fructose elutes around 10-11 minutes and glucoheptose around 16-18 minutes (see figure

2-2 below).

Figure 2-1

Quantitative Fructan Determination Gradient

Table 2-5

Quantitative Fructan Determination Gradient*

Time (min)

Flow

(ml/min)

%C

%D

Curve

0

1.0

10.0

0.0

NA

0

1.0

10.0

0.0

5

20.0

1.0

10.0

0.0

5

20.1

1.0

80.0

20.0

5

30.0

1.0

80.0

20.0

5

30.1

1.0

10.0

0.0

5

45.0

1.0

10.0

0.0

5

*B line not used in this separation, remainder to make up 100% is from line “A” (lab water)

Fos-04 (February 2016)

FOR ERP USE ONLY

DO NOT DISTRIBUTE