AOAC Official methods
–
AOAC 961.16 (Titrimetry)
–
AOAC 944.08 (Ash-Distillation-Titration)
These 2 methods are not suitable for determining low levels of fluoride in infant
formula and adult nutritional products.
–
Microdiffusion and F-ISE (
Microdiffusion and fluoride-specific electrode
determination of fluoride in infant foods: collaborative study
- Dabeka, R. W.,
McKenzie, A. D. J. AOAC., 64, (4): 1021-1026,1981)
Results from collaborative study were not accepted by AOAC. Method not
published.
September 2016
7
German Official Method (Fluo-01)
Amtliche Sammlung von Untersuchungsverfahren, nach Paragraph 35 LMBG.(L 49.00-
7): 2pp., 2000 -
Analysis of foods. Determination of fluoride in dietetic foods with an
ion-sensitive electrode.
•
Only validated on 2 dietetic products
•
Modified and validated with 30 Healthcare products (Nestlé method):
–
PLOQ RTF: ≈50 μg/100g (Dilution factor 2 with TISAB)
–
Good RSDr (< 5%) / RSDiR (< 8%) / Recovery (83%-109%)*
*Bad Recovery for products with high protein/thickeners/caseinates due to non
specific binding of measured free fluorides
Conclusion: method not sensitive & not applicable to products with high protein /
thickener/caseinates content + SLV (SPIFAN kit)
September 2016
8