Ion Chromatography Method (Fluo-02)
ThermoFisher Scientific. Analysis of fluoride in Infant Formula and Adult Nutritionals by
ion chromatography after ultracentrifugation
•
Good linearity (> 0.995) and estimated LOQ (≈ 5μg/100g)
•
Good RSDr (< 5%) , RSDR (<10%) and recovery (81%-105%) for 12 SPIFAN products
•
Specific equipment and IC consumables (column, cartridges, buffer) needed /
Chloride analysed simultaneously
Conclusion: equipment dependent / method to be improved for low F level (e.g.
removal of organic acids, column and buffer used, sample volume injected) for a better
chromatogram interpretation / Complete SLV & additional SLV chloride could be a plus
using improved method
September 2016
9
Microwave acid digestion and F-ISE
10
•
Alternative to tedious & time consuming alkali fusion or microdiffusion
•
LOQ blank: 5 μg/L / LOQsample ≈ 20 μg/100g dry matter (1g with DF=2 TISAB 20%) /
20 food samples (seafood, vegetables, fruits: 0.08 – 1.8 mg/100g dry matter)
•
Not validated for infant formula: (estimated LOQ :< 20 μg/100g RTF)
Conclusion: feasibility to test for Infant Formula with some modifications (e.g. diluted
HNO
3
/ 2g IF reconstituted / F-ISE calibration type)
10
September 2016