Ion Chromatography Method (Fluo-02)

ThermoFisher Scientific. Analysis of fluoride in Infant Formula and Adult Nutritionals by

ion chromatography after ultracentrifugation

Good linearity (> 0.995) and estimated LOQ (≈ 5μg/100g)

Good RSDr (< 5%) , RSDR (<10%) and recovery (81%-105%) for 12 SPIFAN products

Specific equipment and IC consumables (column, cartridges, buffer) needed /

Chloride analysed simultaneously

Conclusion: equipment dependent / method to be improved for low F level (e.g.

removal of organic acids, column and buffer used, sample volume injected) for a better

chromatogram interpretation / Complete SLV & additional SLV chloride could be a plus

using improved method

September 2016

9

Microwave acid digestion and F-ISE

10

Alternative to tedious & time consuming alkali fusion or microdiffusion

LOQ blank: 5 μg/L / LOQsample ≈ 20 μg/100g dry matter (1g with DF=2 TISAB 20%) /

20 food samples (seafood, vegetables, fruits: 0.08 – 1.8 mg/100g dry matter)

Not validated for infant formula: (estimated LOQ :< 20 μg/100g RTF)

Conclusion: feasibility to test for Infant Formula with some modifications (e.g. diluted

HNO

3

/ 2g IF reconstituted / F-ISE calibration type)

10

September 2016

Reference 10 is listed