Methods (FDA)

FDA Method

Spectra and Q/Confirmation Ions of Powdered Kratom Leaf

1200000

1400000

1600000

1800000

2000000

Abundance

Scan 1706 (14.649 min): 07071410.D

397

214

50 100 150 200 250 300 350 400 450 500 550 600 650

0

200000

400000

600000

800000

1000000

m/z-->

269

186

367

75

155

327

129

240 295

103

429 479 553

505

589619647

Challenges

GC Methods

not commonly applied for

M. speciosa

, particularly in the wide 

variety of matrices encountered

High temperatures required to elute the alkaloids – restricts 

parameter adjustment for resolution of any alkaloid mixture

inadequate resolution of mitragynine and speciociliatine

Reliance on mass‐specific detection – EI, ESI, and ESI MS/MS 

spectra of mitragynine, speciogynine, and speciociliatine are 

nearly identical

Interference of speciociliatine – would probably not be detected 

by either DAD or MS detectors

Methods should account for interference