7 . Any general comments about 

the method?

The method is straightforward and clear for all three matrices. For dietary 

supplement the procedure is simple Peaks are well separated in chromatogram

.  

.

This method is lack of separation of beta‐cryptoxanthin isomers and identification of 

lutein and zeaxanthin. 

Beta‐cryptoxanthin standard is included in the method but the purpose isn't clear.

Do you recommend this method be 

adopted as a First Action and

No. This method doesn't demonstrate the capability to separate beta‐cryptoxanthin 

isomers as well as other carotnoids such as lutein and zeaxznthin Thus it doesn't

V.  Recommendation for the Method

published in the Official Methods 

of Analysis of AOAC 

INTERNATIONAL?  Please specify 

rationale.  

,  

meet the requirement of SMPR. In additional, the single laboratory validation result 

was lack of certain data such as reference material, system suitability, LOQ etc.