PDE5-002

ppm.

No time time-to-result information provided.

Overall, not enough information to evaluate the method

performance.

Supporting Data

General Comments

The method was not validated according to the AOAC

2014.012 SMPR for screening of PDE5 inhibitors. Only

method was submitted with minimum method

performance information.

Method Optimization

Evaluate other extraction solvents. Validate the method

according to the AOAC 2014.012 SMPR for screening of

PDE5 inhibitors..

Performance Characteristics

Analytical Range:

No data provided, just information about zone

detectability

LOQ:

Estimated LOD 30 ng/zone (10 ng/μL). Based on the

sample preparation (500 mg sample in 100 mL

methanol), this corresponds to 2 ug/mg (2000 ug/g, 2000

ppm).

Accuracy/Recovery

No data

Precision (RSDr)

No data

Reproducibility (RSDR)

No data

System Suitability

On track 1 of each plate, a mixture of sildenafil,

vardenafil and tadalafil was applied as a system

suitability test.

Safety Review

Is the method safe to use as written?

Yes

Are there any safety concerns with

any component, instrument, reagent

or instruction of the method? If yes,

please clearly state the concerns.

No particular concern.

Does the method include all

necessary safety precautions and/or

warning statements? Yes or No. If

No, please suggest potential

precautions or warnings that can be

added to the method for users or

recommend safer option(s) if

No safety statement is provided. At least a generic

statement should be included, such as:

See AOAC Official Methods of Analysis (2005), Appendix

B: Laboratory Safety. Use appropriate personal

protective equipment such as lab coat, safety glasses,

rubber gloves, and a fume hood. Dispose of solvents and

solutions according to federal, state, and local