MRI Acquisition

Quality Indicator

Importance Rating

(9,essential)

MRI acquisition protocols should generally follow

PIRADSv2 guidelines

(acr.org

) in order to facilitate

diagnostic reporting.

7,7

Replacing coronal and sagittal T2w images with an

isotropic T2w 3D axial acquisition is justified in

radiotherapy planning, especially at 1.5T.

6

When implanted fiducial markers (FMs) or brachytherapy

seeds must be resolved, strategies to augment the FM

signature on above images may suffice (eg. Increase

voxel resolution, PD via dual echo acquisition,

intermediate TE).

7

A separate image may be acquired to more clearly

highlight FMs (e.g. GRE, PD, SSFP, UTE)

7