MRI Acquisition

Quality Indicator

Importance Rating

(9,essential)

DWI is recommended to assist in depiction of tumours if

needed.

9

DCE is considered optional except for salvage of

recurrence.

7

Activation of vendor-provided gradient distortion

correction and optimization of bandwidth (fat/water pixel

shift <1mm) is recommended to maintain geometric

fidelity.

8

DWI should follow the QIBA profile (

qibawiki.rsna.org)

.

Distortion can be minimized using i) R-L phase-encode

direction, ii) volume shimming over the gland, iii)

minimizing effective echo-spacing, and iv) considering

segmented or other non-EPI techniques.

8.5

Motion-related artefacts should be mitigated by limiting

acquisition time to < 5 minutes when possible.

Sequences should be repeated if motion-related artifacts.

7.8