Table 2011.11E. Parameters for UHPLC MS/MS Measurement

Analyte

Q1

a

,

amu

Q3

b

,

amu

Quantifying/Qualifying

Ion

Dwell,

ms

DP

c

EP

d

CE

e

CXP

f

D

3

385.4

259.1

Quantitation Ion

50

55.0

14.5

21.0

17.3

D

3

385.4

107.1

Qualifier Ion

25

55.0

14.5

31.9

5.20

D

3

385.4

159.1

Qualifier Ion

50

55.0

14.5

32.7

9.10

D

2

397.4

125.1

Quantitation Ion

50

43.0

9.5

19.1

6.9

D

2

397.4

107.0

Qualifier Ion

25

43.0

4.5

36.0

19.7

D

2

397.4

271.1

Qualifier Ion

100

43.0

9.0

17.3

18.6

Isotope D

3

388.4

259.1

Internal Standard

50

55.0

14.5

21.0

19.1

Isotope D

2

400.4

125.1

Internal Standard

100

43.0

10

30.0

15.0

a

Q1 = Quadrupole mass filter 1.

b

Q3 = Quadrupole mass filter 3.

c

DP = Declustering potential.

d

EP = Entrance potential.

e

CE = Collision energy.

f

CXP = Collision cell exit potential.

Note: Parameter, except for Mass Spectrometer, Ionization Mode, and Gas Types may be

modified to obtain optimum chromatography and sensitivity. Exact mass ions may vary

slightly from instrument to instrument because of unit resolution of quadrupole mass

spectrometers.