Parameter

Method Performance

Please report

in units as stated in SMPR!

Weighting

factor for

parameter

Suitability Ranking (1-

3-5)

(select from drop-

down list, 5 = best)

SPIFAN matrices

Applicable to all forms of infant, adult, and/or pediatric

formula (powders, ready-to-feed liquids, and liquid

concentrates).

Any combination of milk, soy, rice, whey, hydrolyzed

protein, starch, and amino acids, with and without intact

protein.

All 11 SPIFAN matrices met precision and accuracy

requirements for all nine minerals (plus Cr, Mo, Se)

1

All analytes defined in the

applicability statement are

measured.

Ca, Cu, Fe, Mg, Mn, P, K, Na, Zn

Method determines these nine elements as well as Cr,

Mo, Se simultaneously from same digest.

1

all units mg/100g of RTF

Ca

20 - 1280

same

Cu

0.001 - 1.2

need to verify low end

Fe

0.01 - 20

need to verify low end

Mg

3 - 110

same

Mn

0.001 - 1.0

need to verify low end

P

15 - 800

same

K

10 - 2000

same

Na

10 - 850

same

Zn

0.1 - 18

same

Cr

20-1600 ng/g

same

Mo

20-1000 ng/g

same

Se

10-500 ng/g

same

Ca

20

2.5

Cu

0.001

0.0125

Fe

0.01

0.125

Mg

3.0

0.50

Mn

0.001

0.0125

P

15

1.3

K

10

2.5

Na

10

1.3

Zn

0.1

0.05

Cr

20 ng/g RTF

20

Mo

20 ng/g RTF

20

Se

10 ng/g RTF

10

Ca

90-104, mean 99.6% recovery

Cu

90-106, mean 97.6% recovery

Fe

93-101, mean 97.9% recovery

Mg

93-104, mean 100.5% recovery

Mn

93-101, mean 97.9% recovery

P

98-109, mean 102.9% recovery

K

92-110, mean 103.0% recovery

Na

100-106, mean 103.4% recovery

Zn

93-99, mean 97.4% recovery

Cr

98-102, mean 100% recovery (above PLOQ)

Mo

93-108, mean 95.7% recovery

Se

99-106, mean 102.8% recovery

Ca

1.7% higher than certified

Cu

2.4% lower than certified

Fe

0.9% lower than certified

Mg

1.3% lower than certified

Mn

1.4% lower than certified

P

7.7% higher than certified

K

0.7% higher than certified

Na

0.6% higher than certified

Zn

2.0% higher than certified

Cr

2.3% lower than certified

Mo

3.8% lower than certified

Se

0.6% higher than certified

Cu, Fe, Zn, Mg

5%

method requires duplicate checks <5%

P, Na, Ca, K

5%

method requires duplicate checks <5%

Mn >0.005

5%

method requires duplicate checks <5%

Mn <0.005

8%

method requires duplicate checks <5%

Cr, Mo Se

5%

method requires duplicate checks <5%

Cu, Fe, Zn, Mg

not stated

Cu=1.7-4.9%;Fe=1.9-4.3%;Zn=1.5-2.6%;Mg=1.8-3.4%

P, Na, Ca, K

not stated

P=1.7-5.0%;Na=1.8-3.2%;Ca=1.8-3.5%;K=1.1-3.9%

Mn >0.005

not stated

Mn- 1.3-4.8%

Mn <0.005

not stated

not tested

Cr, Mo, Se

not stated

Cr=1.3-6.3%; Mo=2.6-5.1%; Se = 2.7-6.5%

Cu, Fe, Zn, Mg

10%

Cu=2.3-8.4%;Fe=3.5-5.5%;Zn=2.3-9.7%;Mg=6.2-16.1%

P, Na, Ca, K

8%

P=4.8-

9.7

%; Na=6.5-7.7%;Ca=4.7-6.5%;K=4.5-6.2%

Mn >0.005

10%

Mn = 3.2-5.7%

Mn <0.005

16%

not tested

Cr, Mo, Se

15%

Cr = 4.7-13.0%; Mo = 3.0-7.9%; Se = 2.5-7.3%

b

Units

Bias against established method

Analytical equipment

1

Adequate proof of performance via system suitability

Is there a bias Yes/No ?

Analytical equipment is commonly available in most labs.

a

Concentrations apply to (1) "ready to feed liquids" "as-is"; (2) reconstituted powders (25 g into 200 g water); and (3) liquid concentrates diluted 1:1 by weight.

Did Method Author Consider ERP’s Method Specific

Recommendations

(See web link to specific method

comments):

Decision by ERP

Proprietary equipment

Laboratory safety

Other Considerations

move to Final Action/repeal/remove/expand 2 year term

0

No unique proprietary equipment/accessories are required.

Method does not require any special safety precautions e.g. personal protection from highly toxic solvents.

Evaluation of Method Performance vs. SMPR requirements.

Feedback from Users of the Method since being

awarded First Action

Official Methods

Status

Reproducibility (RSDR)

ADDITIONAL EVALUATION PARAMETERS

SMPR Requirement

AOAC SMPR: 2014.004 (with 2011.009)

Method Reference #

MTE-02 (with additions from Chrom-06)

Method title:

Principle of the method:

Ca, Cu, Fe, Mg, Mn, P, K, Na, Zn by ICP-MS (this method also does Cr, Mo, Se)

Samples are MW digested with an internal standard and analyzed by ICP-MS via external calibration

2

Limit of quantification (LOQ)

Notes:

Overall Score

Recommendation of ERP

2 years after First Action Status

90-110 % mean spiked recovery over the Range

(all elements)

Spike recovery (%)

Bias vs SRM

Use SRM 1849a

2

3

Repeatability (RSDr)

3

Intermediate Reproducibility

(RSDiR)

1

Analytical Range.

1

SPIFAN ERP Checklist v 1.6

27.06.2013

MTE-02 MEF Form