Table 5. Calibration standards and PLOQ for MTE-02

Na

Mg

P

K

Ca

Mn

Fe

Cu

Zn

Cr

Mo

Se

mg/L

mg/L

mg/L

mg/L

mg/L

mg/L

mg/L

mg/L

mg/L

µg/L

µg/L

µg/L

Cal Blank

0

0

0

0

0

0

0

0

0

0

0

0

Cal Std 1

0.50

0.20

0.50

1.0

1.0

0.0050 0.050 0.0050

0.02

0.80

0.80

0.40

Cal Std 2

2.5

1.0

2.5

5.0

5.0

0.025

0.25

0.025

0.10

4.0

4.0

2.0

Cal Std 3

10

4.0

10

20

20

0.1

1.0

0.10

0.40

16

16

8.0

Cal Std 4

20

8.0

20

40

40

0.2

2.0

0.20

0.80

32

32

16

PLOQ

0.25

0.10

0.25

0.50

0.50

0.0025 0.025 0.0025 0.010

0.40

0.40

0.20

50x DF

12.5

5.0

12.5

25

25

0.125

1.25

0.125

0.50 mg/kg RTF

20

20

10 ng/g RTF

The LOQ for this method is the practical LOQ determined from the linearity expt (Table 5a). The PLOQ is half the concentration of the lowest calibration std.

If this PLOQ is multiplied by the typical ICP-MS dilution factor of 50, the results are expressed in terms of mg/kg of an RTF sample.