Table of Contents Table of Contents
Previous Page  212 / 363 Next Page
Information
Show Menu
Previous Page 212 / 363 Next Page
Page Background

Alternative B: Accelerated 74-Hour Extraction at 25–60

°

C

C. Apparatus

(a) Covered water bath capable of maintaining a temperature of up to 60°C for extended

periods. Ensure the mean temperature in the system is 50.0, 55.0, or 60.0 ± 1.0°C by

monitoring incoming and exit temperatures to the manifold at comparable locations. Before

Extraction Nos. 2, 3, and 4 begin, it is necessary to preheat the bath several degrees (

see

X, Y,

and Z in the

Extraction

section below) above the desired temperature to account for initial

heat exchange and temperature equilibration with manifold and columns. Within 10 min, the

bath should be

stabilized at temperature X, Y, or Z.

(b) Reversible peristaltic pump capable of delivering 4.0 (±0.1) mL/min continuously for 54 h.

Pump heads capable of using 16 to 40 tubes are used for an eight to 20 column apparatus,

respectively (Lsmatc ® No. 78006–00 by Cole-Parmer, Vernon Hills, IL).

(c) Extraction apparatus consisting of two parts (illustrated in Figures

2015.15C–F

). Example

equipment as a parts list can be found in Appendix A of ref. 2 in the online

J. AOAC Int

.

(1) Vertical jacketed chromatography columns enclosing inner column of 2.5 × 30 cm (e.g., Ace

Glass, Vineland, NJ; No. 5821–24, filter removed with Teflon ® adapter No. 5838–51). PTFE rods

(6 mm × 15 cm) should be used to avoid channeling of air or caking. All fittings needed to attach

column, pump tubes, and transfer tubing (e.g., 2.4 mm id tubing). Standardize the length of

tubing for each column (typically about 0.75 meter) and secure fittings.

(2) Constant temperature water circulation manifold and pump system capable of maintaining

adequate (minimum 4 L/min) flow and stable temperature for each column. Insulation is typically

required to maintain a stable temperature. Two inline, symmetrically placed thermometers (Figure

2015.15E

) are used tomonitor temperature to input and outflow of manifolds. Attach roll clamps

and flow monitors to column manifold tubing to ensure balanced flows and uniform

temperatures.

(d) Solvent/extract reservoirs [500 mL volumetric flasks (e.g., Kimball Chase Life Science, Vinland,

NJ; No. 28100–500)] with three-hole stoppers and properly placed rigid tubing attached transfer

tubing and to pump (

see

Figure

2015.15E

). Make sure return tube remains approximately 2

cm from the bottom of the flask to prevent pickup of any precipitates.

(e) Detection equipment capable of analyzing liquids at moderate to high (100–10 000 mg/L)

nutrient levels. Analysis falling below the LOD or LOQ should be noted.

(f) 250 mL graduated cylinders (8–20).

D. Reagents and Reference Materials

Candidates for 2016 Method of the Year

211