Table 2015.18B. ICP calibration standards from commercial custom blend stock standard

solutions

Standard

ID

Volume,

mL

Citrate,

mL

Stock

2500 P

,

mL

a

Stock

7500 K,

mL

b

P concn,

µg/mL

P

2

O

5

,

µg/mL

P

2

O

5

solution,

%

P

2

O

5

fertilizer, %

K concn,

µg/mL

K

2

O,

µg/mL

K

2

O

solution, %

K

2

O

fertilizer, %

Blank

1000

400

0

0

0

0

0

0

0

0

0

0

1

250

100

10 of Std 7

NA

12

27.5

0.00275

1.4

360

434

0.04336

21.7

2

250

100

20 of Std 7

NA

24

55

0.00550

2.7

300

361

0.03614

18.1

3

250

100

5

NA

50

115

0.01146

5.7

240

289

0.02891

14.5

4

250

100

10

NA

100

229

0.02291

11.5

180

217

0.02168

10.8

5

250

100

15

NA

150

344

0.03437

17.2

120

144

0.01441

7.2

6

250

100

22

c

NA

220

504

0.05041

25.2

63

76

0.00759

3.8

7

250

100

30

NA

300

687

0.06874

34.4

NA

NA

NA

NA

8

250

100

40

NA

400

917

0.09165

45.8

33.6

40

0.00405

2.0

9

250

100

50

NA

500

1146

0.11457

57.3

16.8

20

0.00202

1.0

10

250

100

NA

25

NA

NA

NA

NA

450

542

0.05421

27.1

11

250

100

NA

30

NA

NA

NA

NA

600

723

0.07227

36.1

12

250

100

NA

35

NA

NA

NA

NA

750

903

0.09034

45.2

13

250

100

NA

35

NA

NA

NA

NA

900

1084

0.10841

54.2

14

250

100

NA

35

NA

NA

NA

NA

1050

1265

0.12648

63.2

a

Stock 1 = 2500 μg/mL P as PO

4

custom stock standard [see Alternative A, B(

p

)] Also, 10 of Std 7

represents a serial dilution from another standard, meaning take 10 mL from standard 7.

b

Stock 2 = 7500 μg/mL K from KCl custom stock standard [see Alternative A, B(

q

)].

c

22 mL can be achieved by using a 15 mL and a 7 mL class A pipet or equivalent combination.