Table of Contents Table of Contents
Previous Page  259 / 363 Next Page
Information
Show Menu
Previous Page 259 / 363 Next Page
Page Background

Table 2015.18C. Calibration criteria for “direct-available” P and “soluble” K by ICP-OES

Element ID

Wavelength, nm

a

Calibration range, μg/mL

Standards used (

see

Table

2015.18A

)Curve fit

Spectral deconvolution

P

177.434 (1)

0 to 100

Blank, 1, 2, 3, 4

Linear

None

P

177.434 (2)

100 to 500

4, 5, 6, 7, 8, 9

Linear

None

P

178.222 (1)

0 to 100

Blank, 1, 2, 3, 4

Linear

None

P

178.222 (2)

100 to 500

4, 5, 6, 7, 8, 9

Linear

None

P

213.618 (1)

0 to 100

Blank, 1, 2, 3, 4

Linear

Cu 213.598

P

213.618 (2)

100 to 500

4, 5, 6, 7, 8, 9

Linear

Cu 213.598

P

214.914 (1)

0 to 100

Blank, 1, 2, 3, 4

Linear

Cu 214.898

P

214.914 (2)

100 to 500

4, 5, 6, 7, 8, 9

Linear

Cu 214.898

K

766.485 (1)

0 to 126

Blank, 1, 2, 3, 4

Quadratic

None

K

766.485 (2)

50 to 379

3, 4, 5, 6, 7

Quadratic

None

K

769.897

126 to 505

4, 5, 6, 7, 8

Quadratic

Interference LiNO

3

K

404.721

505 to 1046

8, 9, 10, 11, 12

Quadratic

None

a

(1) and (2) distinguish between the same wavelength used to cover two separate concentration ranges.

Table 2015.18D. Final ICP-OES conditions utilized for citrate-EDTA-soluble P and K validation

Factor

Setting

Power, kW

1.45

Plasma flow, L/min

19.5

Auxiliary flow, L/min

2.25

Nebulizer pressure, L/min

0.7

Nebulizer type

Seaspray

Spray chamber

Cyclonic

Sample pump tube

Black/black

Buffer/internal standard pump tube

Gray/gray

CsCl ionic buffer concn, M

0.018

Internal standard and concn, µg/mL

10

Buffer matrix

4% nitric acid

Exposure length, s

10

Number of exposures

3

Rinse time, s

35

Total analysis time, min

2

Candidates for 2016 Method of the Year

258