![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0259.png)
Table 2015.18C. Calibration criteria for “direct-available” P and “soluble” K by ICP-OES
Element ID
Wavelength, nm
a
Calibration range, μg/mL
Standards used (
see
Table
2015.18A
)Curve fit
Spectral deconvolution
P
177.434 (1)
0 to 100
Blank, 1, 2, 3, 4
Linear
None
P
177.434 (2)
100 to 500
4, 5, 6, 7, 8, 9
Linear
None
P
178.222 (1)
0 to 100
Blank, 1, 2, 3, 4
Linear
None
P
178.222 (2)
100 to 500
4, 5, 6, 7, 8, 9
Linear
None
P
213.618 (1)
0 to 100
Blank, 1, 2, 3, 4
Linear
Cu 213.598
P
213.618 (2)
100 to 500
4, 5, 6, 7, 8, 9
Linear
Cu 213.598
P
214.914 (1)
0 to 100
Blank, 1, 2, 3, 4
Linear
Cu 214.898
P
214.914 (2)
100 to 500
4, 5, 6, 7, 8, 9
Linear
Cu 214.898
K
766.485 (1)
0 to 126
Blank, 1, 2, 3, 4
Quadratic
None
K
766.485 (2)
50 to 379
3, 4, 5, 6, 7
Quadratic
None
K
769.897
126 to 505
4, 5, 6, 7, 8
Quadratic
Interference LiNO
3
K
404.721
505 to 1046
8, 9, 10, 11, 12
Quadratic
None
a
(1) and (2) distinguish between the same wavelength used to cover two separate concentration ranges.
Table 2015.18D. Final ICP-OES conditions utilized for citrate-EDTA-soluble P and K validation
Factor
Setting
Power, kW
1.45
Plasma flow, L/min
19.5
Auxiliary flow, L/min
2.25
Nebulizer pressure, L/min
0.7
Nebulizer type
Seaspray
Spray chamber
Cyclonic
Sample pump tube
Black/black
Buffer/internal standard pump tube
Gray/gray
CsCl ionic buffer concn, M
0.018
Internal standard and concn, µg/mL
10
Buffer matrix
4% nitric acid
Exposure length, s
10
Number of exposures
3
Rinse time, s
35
Total analysis time, min
2
Candidates for 2016 Method of the Year