Table of Contents Table of Contents
Previous Page  61 / 199 Next Page
Information
Show Menu
Previous Page 61 / 199 Next Page
Page Background

B

ird

et al

.:

J

ournal of

AOAC I

nternational

V

ol

.

99, N

o

.

3, 2016 

671

interlaboratory data analyses are presented in Tables

2015.13A

and

2015.13B

.

Raw Easy-Peel Shrimp

Raw easy-peel shrimp test portions were evaluated at a

low, medium, and high contamination level for the presence of

naturally occurring aerobic bacteria. Replicate 3M Petrifilm RAC

Plates were evaluated at both 32 and 35°C (Tables

2015.13D

and

 E

) Sixteen laboratories participated in the analysis of this

matrix, with all 16 laboratories submitting data.

3M Petrifilm RAC Plate Incubated at 32°C

For the 3M Petrifilm RAC Plate low and medium

contamination levels and for the FDA BAM low and high

contamination levels, one outlier laboratory was identified

by the Cochran’s tests for outliers. However, no evidence of

physical cause or suspicion of cause was noted and it was

determined that the data would be included in the statistical

analysis.

There were no statistically significant differences between the

3M Petrifilm RAC Plate and FDABAM methods as determined

by the 95% CI of the differences of means at any of the three

contamination levels.

3M Petrifilm RAC Plate Incubated at 35°C

For the 3M Petrifilm RAC Plate medium contamination level

and for the FDA BAM low and high contamination levels, one

outlier laboratory was identified by the Cochran’s tests for

outliers. However, no evidence of physical cause or suspicion

of cause was noted and it was determined that the data would be

included in the statistical analysis.

Figure 3. Youden plots for 3M Petrifilm RAC Plate and SMEDP results for pasteurized skim milk.

Figure 2. Youden plots for 3M Petrifilm RAC Plate and FDA BAM results for raw easy-peel shrimp evaluated at 35°C.