There were no statistically significant differences between the

3M Petrifilm RAC Plate and FDABAM methods as determined

by the 95% CI of the differences of means at any of the three

contamination levels.

Pasteurized Skim Milk

Pasteurized skim milk test portions were evaluated at a

low, medium, and high contamination level for the presence

of naturally occurring aerobic bacteria. Fifteen laboratories

participated in the analysis of this matrix, with 13 laboratories

submitting data.

For the SMEDP high contamination level, one outlier

laboratory was identified by the Cochran’s tests for outliers.

However, no evidence of physical cause or suspicion of cause

was noted and it was determined that the data would be included

in the statistical analysis.

There were no statistically significant differences between the

3M Petrifilm RAC Plate and SMEDP methods as determined

by the 95% CI of the differences of means at any of the three

contamination levels.

Instant NFDM

Instant NFDM test portions were evaluated at a low, medium,

and high contamination level for the presence of naturally

occurring aerobic bacteria. Fifteen laboratories participated in

the analysis of this matrix, with 15 laboratories submitting data.

For the 3M Petrifilm RAC Plate low contamination level

and for the SMEDP medium contamination level, one outlier

laboratory was identified by the Cochran’s tests for outliers.

Figure 4. Youden plots for 3M Petrifilm RAC Plate and SMEDP results for instant NFDM.

Figure 5. Youden plot for combined mean 3M Petrifilm RAC Plate and FDA BAM results for raw easy-peel shrimp evaluated at 32°C.