Single Laboratory Validation Report for Total Amino Acids by UHPLC-UV in Infant Formulas and Adult Nutritionals

Jaudzems, Lahrichi, and Fuerer

January 2017

page 15/20

Tyr

Val

From the correlation coefficient R

2

of the regression lines (> 0.99), the standard deviation of

the residuals, and the graphs of the residuals, we conclude that the calibration curves are

linear between

(0.25 and 10) pmol/μL for Cys2 and (

0.25 and 50) pmol/µL for all other amino

acids.

4.3

Limit of quantification

The scope of this method implies working at effective concentrations higher than the LoQ

stipulated in the SMPR. Therefore, the LoQ for the different amino acids were set to the

lowest point of their linearity ranges (see point

4.2 above)

. This is equivalent to 0.25 pmol per

injection (1 µL) for all amino acids studied.

Using an average molecular weight of 110 g/mol and a sample mass of 200 mg

(reconstituted powder),

0.25 pmol/μL (derivatized sample) corresponds to 2.5 mg/100 g in

the reconstituted sample according to the following three equations:

Equation 1: concentration of analyte in the hydrolysate (4 and 10 are the dilutions occurring

during the neutralization and derivatization steps, respectively).

Equation 2: amount of analyte in the sample

Equation 3: amount of analyte in 100 g of reconstituted powder