Single Laboratory Validation Report for Total Amino Acids by UHPLC-UV in Infant Formulas and Adult Nutritionals

Jaudzems, Lahrichi, and Fuerer

January 2017

page 19/20

Two products show strong matrix effects with bias towards either low (#4) or high (#12)

recoveries.

Finally, recoveries for Product 6 (NIST 1849A) were compared with the reference values

from the certificate of analysis (CoA).

4.10

Chromatograms

Example chromatograms for selected concentrations of the calibration curve and SPIFAN II

SLV kit products are presented in Appendix I.

Analyte

6

(spike)

6

(CoA)

His

92.6% 99.7%

Tau

97.0% 100.2%

Ser

104.0% 98.0%

Arg

94.8% 97.4%

Gly

97.9% 100.7%

Asp

105.2% 106.2%

Glu

103.4%

106.1%

Thr

99.6%

124.1%

Ala

99.5% 103.2%

Pro

100.2% 97.7%

Lys

100.6%

107.7%

Tyr

96.1%

113.4%

xCys

99.4% 102.4%

Met

95.4% N/A

Val

91.6% 107.5%

Ile

89.8% 108.7%

Leu

96.8% 99.9%

Phe

96.8% 101.6%

AMINO-01 (FEBRUARY 2017)

SLV REPORT

FOR ERP USE ONLY

DO NOT DISTRIBUTE