FOR EXPERT REVIEW PANEL USE ONLY
NOT FOR DISTRIBUTION
2
Cons/Weaknesses
•
Supporting Data
•
General Comment:
Concerned over only 9 labs participating in MLT.
-
Method Optimization:
•
Performance Characteristics:
Analytical Range:
0.1 to 300 which exceeds SMPR of 0.1 to 150
LOQ:
Meets LOQ of 0.1 mcg/100g
Accuracy/Recovery:
Recoveries 95 to 105%. Meets SMPR of 90 to 110
Precision (RSD
r
): N/A
Reproducibility (RSD
R
):
All RSDs are <10% meeting SMPR
•
System suitability: