so the longer the cables the bigger

the error.

The result would be considered to

be excellent even if the result were

further from the nominal, in order

to account for these additional

tolerances. It is clear that the Picotest

J2102A common mode transformer

and a preamp are indispensable when

it comes to getting the most accuracy

during PDN measurement.

Figure 5 Impedance measurement of the 2mΩ resistor, using Port 1

and Port 2, with and without the common mode coaxial transformer

on the Keysight E5061B. The unique GP ports don’t require the

transformer, but are limited to 30MHz

Figure 5 A DC measurement

is made using precision, low

noise 6.5 Digit DMMs (Picotest

M3500A) to obtain a second

measurement for correlation.

The power supply does not have

current programming, which is

the reason the current is set to an

odd number. The two channels of

the power supply are connected

in parallel and both are in current

limit for this measurement.

Figure 6 - Noise floor in Ohms on

the Bode 100 using the J2102A

and the J2180A preamplifier

included

Figure 4, The Bode 100

measurement of the 250uOhm

resistor shown in Figure 1. The

measured resistance and ESL are

shown

New-Tech Magazine Europe l 61