SUBMITTED 050113

Operator 2 – Test each of the 84 PBS resuspensions in the ANSR

Salmonella

assay following the

instructions in the section “For Use as a Confirmation Test from Colony Picks” in the ANSR

Salmonella

kit insert. There are 16 positions on the ANSR reader. Samples should be tested in

groups of 15, with the 16

th

assay position reserved for a PBS blank (0.5 mL PBS, treated as a test

sample) in each experimental run. Operator 1 is not to participate in or observe this activity.

Run 1 Samples 1-15 plus PBS blank

Run 2 Samples 16-30 plus PBS blank

Run 3 Samples 31-45 plus PBS blank

Run 4 Samples 46-60 plus PBS blank

Run 5 Samples 61-75 plus PBS blank

Run 6 Samples 76-84 plus PBS blank

Each experimental run will require approximately 30 minutes.

Record test results on the Data Recording Form (appended). Record results as “Positive” or

“Negative”. If any samples produce a result of “Invalid”, record this result, then repeat the test

using the same PBS resuspension and record the second result alongside the initial result.

Day 5 Transmit Data Recording Forms and ANSR assay raw data (.json files) to the Study Director.

Instructions will be provided.