D. (c)

After incubation, homogenize samples manually. Follow appropriate instructions

1

based on heating method.

2

3

Boiling

- Transfer 2-3 mL of the enrichment broth into a tube. Seal the tube. Heat

4

in a water bath for 5± 1 minutes at 95-100°C. Cool the tube. Mix the boiled

5

broth and transfer 0.5 mL into the sample well of the VIDAS LPT reagent strip.

6

Perform the VIDAS test.

7

8

Heat and Go

- Transfer 0.5 mL of the enrichment broth into the sample well of

9

the VIDAS LPT reagent strip. Heat for 5± 1 minutes (see VIDAS Heat and Go

10

User’s Manual). Remove the strip and allow to cool for 10 minutes prior to test

11

initiation. Perform the VIDAS test.

12

13

14

E. Enzyme Immunoassay

15

E. (a)

Enter factory master calibration curve data into the instrument using the master lot

16

entry (MLE) card.

17

E. (b)

Remove the kit reagents and materials from refrigerated storage and allow them to

18

come to room temperature for at least 30 minutes.

19

E. (c)

Use one VIDAS LPT Reagent Strip and one VIDAS LPT SPR for each sample,

20

control or standard to be tested. Reseal the storage pouch after removing the required

21

number of SPRs.

22

E. (d)

Enter the appropriate assay information to create a work list. Enter the test code

23

by typing or selecting “LPT”, and number of tests to be run. If the standard is to be

24

tested, identify the standard by “S1” and test in duplicate. If the positive control is to be

25

tested, identify it by “C1”. If the negative control is to be tested, identify it by “C2”.

26

a.

Note: The standard must be tested upon receipt of a new lot of reagents and then

27

every 14 days. The relative fluorescence value (RFV) of the standard must fall

28

within the set range provided with the kit.

29

E. (e)

Load the LPT reagents strips and SPRs into the positions that correspond to the

30

VIDAS section indicated by the work list. Verify that the color labels with the assay code

31

on the SPRs and reagent strips match.

32

E. (f)

Initiate the assay processing as directed in the VIDAS operator’s manual.

33

E. (g)

After the assay is completed, remove the SPRs and reagent strips from the

34

instrument and dispose of properly.

35

36

F. Results and Interpretation

37

The results are analyzed automatically by the VIDAS system. A report is printed which

38

records the type of test performed, the test sample identification, the date and time, the lot

39

number and expiration date of the reagent kit being used, and each sample’s relative

40

fluorescence value (RFV), test value, and interpreted result (positive or negative).

41

Fluorescence is measured twice in the Reagent Strip’s reading cuvette for each sample

42

tested. The first reading is a background reading of the substrate cuvette before the SPR

43

is introduced into the substrate. The second reading is taken after incubating the substrate

44

8