Table of Contents Table of Contents
Previous Page  964 / 1195 Next Page
Information
Show Menu
Previous Page 964 / 1195 Next Page
Page Background

Eurofins IPL Nord SAS – 1 rue du professeur Calmette – 59046 LILLE cedex

9 / 15

VIDAS LMX – Summary 2012 v01

3.2 Control of experimental parameters

3.2.1 Contamination levels obtained after artificial contamination

The following table shows the contamination rates obtained and estimated precisions :

Level

Samples

Targeted theoretical

rate (CFU/25ml)

Real rate

(CFU/25ml)

Estimated lower

contamination limit

per 25 ml sample

Estimated upper

contamination limit

per 25 ml sample

Level 0 (L0)

5-6-9-10-13-14-19-20

27-28-35-36-41-42-43-44

0

0

/

/

Low level (L1)

3-4-11-12-17-18-23-24

29-30-31-32-39-40-45-46

3

3,26

0,7

9,5

High level (L2)

1-2-7-8-15-16-21-22

25-26-33-34-37-38-47-48

30

22,2

14

33

3.2.2 Problems of temperature recorded during transport, temperature on reception and

reception times

a)

Analysis of temperature monitoring curves during transport

Temperatures registered by thermo button during shipment were stables and comprised between 0°C and 8° C for

all laboratories.

b)

Temperatures on reception and reception times

The temperatures obtained are recorded in the following table :

Laboratory

Reception Temperatures (°C)

Comments

communicated by the laboratory

indicated by the thermo button

A

3.1

1.2

B

4.5

2.5

C

3.3

1.5

D

4.2

4.0

E

2.7

3.0

F

Not communicated

2.5

Withdrawal

G

Not communicated

1.5

H

2.6

3.0

I

6.0

3.5

J

4.1

4.0

Reception at D1, assays at D2

K

6.1

0.1

L

4.8

-0.4

M

1.4

2.1

N

3.9

3.5

O

7.0

/

Reception at D2

P

6.0

4.1

Reception at D2

Among the16 laboratories, 14 laboratories received samples the day after the sending.

3.2.3 Conclusion

Continuation in the conditions of transport of samples, the results are exploitable for 12 laboratories (exclusion from

laboratories C, F, J, I, L and P).

Further to the sending of samples in 16 laboratories, the results are exploitable for 12 laboratories (exclusion from

laboratories J, O and P continuation in the conditions of reception of samples at D2 and withdrawal of the

laboratory F).