Silliker

Chem, Res. Center Crete, IL – Report of A Validation of LC-MS/MS Method for Folate Analysis

15

13

C

5

-FA

447.2

295.2

20

4.5

39

31

6

13

C

5

-5-CH3-

THF

465.2

313.2

35

5

23.9

29

6

DP: Declustering potential;

EP Entrance potential;

CEP collision cell entrance potential;

CE Collision cell;

CXP collision cell exit potential.

The General MS parameters for API 5500 (Qtrap):

Curtain gas (nitrogen):

25 psi;

Collision Assisted Dissociation: Medium

Ion Spray voltage:

5500 V;

Temperature:

700 °C;

Ion source gas1 (zero air):

60 psi;

Ion source gas2 (zero air):

50 psi;

Calculations :

Calculations and reporting

Ratio of the folate vitamer with respective internal standard is determined by

HPLC-MS analysis. 5-CH3-THF, 5-CHO-THF and THF use labeled 5-CH3-THF as

internal std.

Folic acid and 10-CH3-Folic acid 10-CHO-Folic acid use labeled Folic acid as internal

std. Calibration curve is drawn based on concentration of calibration standard of

specific folate compound plotted against ratio of peak area (for the analyte) to the

relevant internal standard. The calibrations curve are prepared for all the targeted

folate compounds. The ratio peak areas of the folate compounds and respective

internal standard is calculated in the standard sample which was processed like

sample. Concentration of each of the folate compound as ng/mL in the standard

processed like sample is calculated based on calibration curve. This is usually

determined by the MS software (multquant). The ratio of peak areas of the

detected folate compounds and respective internal standard is calculated in the

blank which was processed like sample in the batch. Concentration of each of the

detected folate compound as ng/mL is calculated in blank based on calibration

curve. . Concentration of each of the folate compound as ng/mL in the standard

processed like sample is adjusted by substracting the amount of respective folate

compound detected in blank. Recovery factors (Rfs) for all the targeted folate

compounds are determined based on adjusted concentration of respective analyte

determined in standard processed like sample using the following relationship.

Rf = Adjusted concentration (ng/mL) of folate compound estimated in std.

/amount of the specific folate compound expected to be present in the std.

The ratio peak areas of the folate compounds and respective internal standard is

calculated in a sample. Concentration of each of the folate compound as ng/mL

2011.06 (Fol-22) w/SLV