Chem, Res. Center Crete, IL – Report of A Validation of LC-MS/MS Method for Folate Analysis
19
method.
AOAC method uses formic Acid. We explored the use of formic acid and acetic acid at various
concentrations and pH’s. The use of 1%Acetic Acid (approx. pH 2.8) gave the best response.
Table 12below shows comparison between Formic and Acetic.
Table 12 : Analytical Response of different analytes with different mobile phase buffers
Mobile Phase
FA
5-CH3-
THF
5-CHO-
THF
THF
10-CH3-
FA
10-CHO-
FA
10 mM
NH4Frmt
pH=2.8
2.22E+05
6.00E+05
6.54E+05
2.42E+05
3.96E+05
6.15E+05
2.14E+05
6.07E+05
6.35E+05
2.61E+05
4.01E+05
6.00E+05
2.18E+05
6.01E+05
6.32E+05
2.55E+05
3.62E+05
5.83E+05
Average
2.18E+05
6.03E+05
6.40E+05
2.53E+05
3.86E+05
5.99E+05
%RSD
1.8%
0.6%
1.9%
3.8%
5.5%
2.7%
1% Acetic
Acid
3.79E+05
1.67E+06
1.39E+06
5.87E+05
5.84E+05
1.38E+06
3.67E+05
1.67E+06
1.35E+06
5.89E+05
6.29E+05
1.36E+06
4.15E+05
1.69E+06
1.31E+06
6.24E+05
5.83E+05
1.35E+06
Average
3.87E+05
1.68E+06
1.35E+06
6.00E+05
5.99E+05
1.36E+06
%RSD
6.5%
0.7%
3.0%
3.5%
4.4%
1.1%
Evaluation of different solvents in the mobile phase to get higher analytical response of the analytes in
the method:
The analytical response of the targeted analytes was determined with different solvents as a part of the
mobile phase. The following solvents were evaluated for the stated purpose.
Methanol
Acetonitrile
AOAC method states the use of methanol as Mobile Phase B. Both solvents provide a good
chromatography and a satisfactory resolution of the analytes. We determined the analytical response of
the targeted analytes at 50 ppb level with either Methanol or Acetonitrile as a part of the mobile phase.
Table 13below shows comparison of peak areas when either one of these solvents were used as a part
of the mobile phase.
The analytes which are expected at lower concentrations in food samples , 5-methyl-THF, 5-formyl-
THF and 10-methyly folic acid provided a higher analytical response with Methanol in the mobile
phase in comparison when it was replaced by acetonitrile. Analytical response of other analytes was
more or less same with both solvents.