![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0245.jpg)
S232
ESTRO 36
_______________________________________________________________________________________________
Results
The average beam-on delivery time was 3min, ranging
from 1.22min to 8.82min. The average
ɣ
(3%,3mm) passing
rate for film measurements was 97.0 ±1.6% (range from
93.3 to 98.8%), while the Delta
4
measurements presented
an average
ɣ
(2%,2mm) of 97.7±1.4% (range from 95.3 to
99.6%) respectively. The average isocentre point dose
ratio was 99.9±1.2% (range from 98.0 to 102.8%). For the
lung SBRT verifications with the CIRS phantom, an average
local
ɣ
of 97.0±1,0% and 93.1±2.0% was obtained during
the coronal and sagittal film analysis, whereas the average
isocentre point dose ratio was 100.0±1.4%. An overall
mean gantry-ring geometric deviation of 0.04° ± 0.46° and
0.19° ± 0.26° was obtained, respectively.
Conclusion
DWA has been successfully added to the non-coplanar
rotational IMRT techniques’ arsenal, allowing additional
flexibility in dose shaping while preserving dosimetrically
robust delivery. In a short period of time, it has become a
standard treatment technique on the Vero system in our
department.
OC-0440 Characterization and clinical evaluation of a
novel CT reconstruction to derive electron densities
B. Van der Heyden
1
, M. Ollers
1
, C. Loon Ong
1
, F.
Verhaegen
1
, W. Van Elmpt
1
1
School for Oncology and Developmental Biology-
Maastricht University Medical Centre, Department of
Radiation Oncology MAASTRO- GROW, Maastricht, The
Netherlands
Purpose or Objective
Radiotherapy dose calculations are almost exclusively
performed on CT images. In a clinical workflow, the
Hounsfield Units (HU) are converted into electron density
(ED) by using a CT to ED conversion curve calibrated for a
typically fixed tube potential (e.g. 120 kV). Recently, a
novel CT image reconstruction algorithm (DirectDensity
TM
,
Siemens Healthcare GmbH, Germany) was developed that
directly reconstructs the ED, independent of the tube
potential of the CT scanner. This allows the elimination of
a conversion curve for each tube potential. Our study
describes the accuracy in terms of dose calculation of the
reconstruction
algorithm
based
on
phantom
measurements and shows the application in a clinical
radiation therapy workflow for different tube potentials.
Material and Methods
The accuracy of the novel reconstruction algorithm to
derive ED was validated in a Gammex RMI 467 phantom
(Gammex, USA) using different tissue mimicking inserts.
The phantom was scanned at different tube potentials (80
kV, 120 kV and 140 kV) with a novel SOMATOM
Confidence® RT Pro scanner (Siemens Healthcare GmbH,
Germany). Images were reconstructed both into HU and
ED for each tube potential. Next, the usability of the
reconstruction algorithm was evaluated in a clinical
workflow. Five patients with an abdominal lesion (e.g.
rectal or prostate cancer) were scanned using the
clinically used tube potential of 120 kV and an additional
dual-spiral dual-energy CT acquisition was made at 80 kV
and 140 kV. Dose distributions (Eclipse
TM
, Varian, USA) of
the ED images of the 80 kV, 120 kV, 140 kV acquisitions
using the novel reconstruction algorithm were then
compared with the clinical plan based on the 120 kV
acquisition using the clinical CT to ED curve with the
standard HU image of the 120 kV scan. The difference in
mean doses delivered to the planning target volume were
quantified (i.e. relative difference ± 1 SD).
Results
The CT to ED conversion curve for the HU images
depended on the tube potential of the CT scanner. The
novel reconstruction algorithm produced ED values that
had a residual from the identity line of -0.1% ± 2.2% for all
inserts and energies and is shown in Figure 1.
The dose distributions between the standard and the novel
reconstruction algorithm were compared for different
energies. The relative differences in target dose ranges
were small and ranged from -0.2% to 0.7% for 80 kV, -0.1%
to 1.1% for 120 kV, and 0.1 to 1.0% for 140 kV.
Figure 1: The linear conversion curve (fitted) of the
novel reconstruction algorithm.
Conclusion
A novel reconstruction algorithm to derive directly
relative electron density irrespective of the tube potential
of the CT scanner was evaluated. A single identity curve
for the CT to ED could be used in the treatment planning
system. This reconstruction algorithm may enhance the
clinical workflow by selecting an optimal tube potential
for the individual patient examination that is not
restricted to the commonly used 120 kV tube potential.
OC-0441 Dose Prescription Function from Tumor
Voxel Dose Response for Adaptive Dose Painting by
Number
D. Yan
1
, S. Chen
2
, G. Wilson
1
, P. Chen
1
, D. Krauss
1
1
Beaumont Health System, Radiation Oncology, Royal
Oak MI, USA
2
Beaumont Health System, Radiation Oncology, Royal
Oak, USA
Purpose or Objective
Dose-painting-by-number (DPbN) needs a novel Dose
Prescription Function (DPF) which provides the optimal
clinical dose to each tumor voxel based on its own dose
response. To obtain the DPF for adaptive DPbN, a voxel-
by-voxel tumor dose response matrix needs to be
constructed during the early treatment course. The study
demonstrated that the voxel-by-voxel tumor dose